Impact of an advanced image-based monoenergetic reconstruction algorithm on coronary stent visualization using third generation dual-source dual-energy CT: a phantom study.
Mangold, Stefanie; Cannaó, Paola M; Schoepf, U Joseph; Wichmann, Julian L; Canstein, Christian; Fuller, Stephen R; Muscogiuri, Giuseppe; Varga-Szemes, Akos; Nikolaou, Konstantin; De Cecco, Carlo N.
Afiliação
  • Mangold S; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Ashley River Tower, 25 Courtenay Drive, Charleston, SC, 29425-2260, USA.
  • Cannaó PM; Department of Diagnostic and Interventional Radiology, Eberhard-Karls University Tuebingen, Tuebingen, Germany.
  • Schoepf UJ; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Ashley River Tower, 25 Courtenay Drive, Charleston, SC, 29425-2260, USA.
  • Wichmann JL; Scuola di Specializzazione in Radiodiagnostica, University of Milan, Milan, Italy.
  • Canstein C; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Ashley River Tower, 25 Courtenay Drive, Charleston, SC, 29425-2260, USA. schoepf@musc.edu.
  • Fuller SR; Division of Cardiology, Department of Medicine, Medical University of South Carolina, Charleston, SC, USA. schoepf@musc.edu.
  • Muscogiuri G; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Ashley River Tower, 25 Courtenay Drive, Charleston, SC, 29425-2260, USA.
  • Varga-Szemes A; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany.
  • Nikolaou K; Siemens Medical Solutions, Malvern, PA, USA.
  • De Cecco CN; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Ashley River Tower, 25 Courtenay Drive, Charleston, SC, 29425-2260, USA.
Eur Radiol ; 26(6): 1871-8, 2016 Jun.
Article em En | MEDLINE | ID: mdl-26373752
ABSTRACT

PURPOSE:

To evaluate the impact of an advanced monoenergetic (ME) reconstruction algorithm on CT coronary stent imaging in a phantom model. MATERIALS AND

METHODS:

Three stents with lumen diameters of 2.25, 3.0 and 3.5 mm were examined with a third-generation dual-source dual-energy CT (DECT). Tube potential was set at 90/Sn150 kV for DE and 70, 90 or 120 kV for single-energy (SE) acquisitions and advanced modelled iterative reconstruction was used. Overall, 23 reconstructions were evaluated for each stent including three SE acquisitions and ten advanced and standard ME images with virtual photon energies from 40 to 130 keV, respectively. In-stent luminal diameter was measured and compared to nominal lumen diameter to determine stent lumen visibility. Contrast-to-noise ratio was calculated.

RESULTS:

Advanced ME reconstructions substantially increased lumen visibility in comparison to SE for stents ≤3 mm. 130 keV images produced the best mean lumen visibility 86 % for the 2.25 mm stent (82 % for standard ME and 64 % for SE) and 82 % for the 3.0 mm stent (77 % for standard ME and 69 % for SE). Mean DLP for SE 120 kV and DE acquisitions were 114.4 ± 9.8 and 58.9 ± 2.2 mGy × cm, respectively.

CONCLUSION:

DECT with advanced ME reconstructions improves the in-lumen visibility of small stents in comparison with standard ME and SE imaging. KEY POINTS • An advanced image-based monoenergetic reconstruction algorithm improves lumen visualization in stents ≤3.0 mm. • Application of high keV reconstructions significantly improves in-stent lumen visualization. • DECT acquisition resulted in 49 % radiation dose reduction compared with 120 kV SE.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Doença da Artéria Coronariana / Prótese Vascular / Tomografia Computadorizada por Raios X / Imagem Radiográfica a Partir de Emissão de Duplo Fóton / Stents / Imagens de Fantasmas / Vasos Coronários Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Doença da Artéria Coronariana / Prótese Vascular / Tomografia Computadorizada por Raios X / Imagem Radiográfica a Partir de Emissão de Duplo Fóton / Stents / Imagens de Fantasmas / Vasos Coronários Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2016 Tipo de documento: Article