[Simultaneous quantitative analysis of different ceramide species in cells by high-performance liquid chromatography-tandem mass spectrometry].
Zhejiang Da Xue Xue Bao Yi Xue Ban
; 44(4): 429-34, 2015 07.
Article
em Zh
| MEDLINE
| ID: mdl-26555422
OBJECTIVE: To establish a high-performance liquid chromatography-tandem mass spectrometry (HPLC-ESI-MS) method for simultaneous quantitative analysis of different ceramide species in cells. METHODS: The analysis was performed on an Agilent 1290 HPLC system with a ZORBAX Eclipse XDB-C8 (150.0 mm × 2.1 mm, 3.5 µL) column and a temperature of 35 â. Methanol with 1 mmol/L ammonium formate and 0.2% formic acid was used as mobile phase A and 100% methanol was used as mobile phase B. And the ceramides were separated by gradient elution at a flow rate of 0.3 mL/min. Electrospray ionization (ESI) with multiple reaction monitoring (MRM) was used in the analysis. RESULTS: Four ceramide species all had a good linear response in the determination ranges (R² ≥ 0.9987). The average recoveries (n = 9) were 99.1%,99.9%,100.5% and 98.2% with RSDs of 5.6%, 5.1%, 4.7% and 5.5%, respectively. In addition, the levels of ceramides in FL cells were relatively stable, while the C24-ceramide had the highest level. CONCLUSION: The HPLC-ESI-MS method for simultaneous analysis of ceramides has high accuracy, reproducibility and linearity, which may be used for quantification of ceramide species in cells.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Ceramidas
Idioma:
Zh
Ano de publicação:
2015
Tipo de documento:
Article