Your browser doesn't support javascript.
loading
Real-time Monitoring of High Intensity Focused Ultrasound (HIFU) Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound (HMIFU).
Grondin, Julien; Payen, Thomas; Wang, Shutao; Konofagou, Elisa E.
Afiliação
  • Grondin J; Department of Biomedical Engineering, Columbia University.
  • Payen T; Department of Biomedical Engineering, Columbia University.
  • Wang S; Department of Biomedical Engineering, Columbia University.
  • Konofagou EE; Department of Biomedical Engineering, Columbia University; Department of Radiology, Columbia University; ek2191@columbia.edu.
J Vis Exp ; (105): e53050, 2015 Nov 03.
Article em En | MEDLINE | ID: mdl-26556647
ABSTRACT
Harmonic Motion Imaging for Focused Ultrasound (HMIFU) is a technique that can perform and monitor high-intensity focused ultrasound (HIFU) ablation. An oscillatory motion is generated at the focus of a 93-element and 4.5 MHz center frequency HIFU transducer by applying a 25 Hz amplitude-modulated signal using a function generator. A 64-element and 2.5 MHz imaging transducer with 68kPa peak pressure is confocally placed at the center of the HIFU transducer to acquire the radio-frequency (RF) channel data. In this protocol, real-time monitoring of thermal ablation using HIFU with an acoustic power of 7 W on canine livers in vitro is described. HIFU treatment is applied on the tissue during 2 min and the ablated region is imaged in real-time using diverging or plane wave imaging up to 1,000 frames/second. The matrix of RF channel data is multiplied by a sparse matrix for image reconstruction. The reconstructed field of view is of 90° for diverging wave and 20 mm for plane wave imaging and the data are sampled at 80 MHz. The reconstruction is performed on a Graphical Processing Unit (GPU) in order to image in real-time at a 4.5 display frame rate. 1-D normalized cross-correlation of the reconstructed RF data is used to estimate axial displacements in the focal region. The magnitude of the peak-to-peak displacement at the focal depth decreases during the thermal ablation which denotes stiffening of the tissue due to the formation of a lesion. The displacement signal-to-noise ratio (SNRd) at the focal area for plane wave was 1.4 times higher than for diverging wave showing that plane wave imaging appears to produce better displacement maps quality for HMIFU than diverging wave imaging.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ablação por Ultrassom Focalizado de Alta Intensidade / Fígado Tipo de estudo: Guideline Limite: Animals Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ablação por Ultrassom Focalizado de Alta Intensidade / Fígado Tipo de estudo: Guideline Limite: Animals Idioma: En Ano de publicação: 2015 Tipo de documento: Article