Your browser doesn't support javascript.
loading
Slow light Mach-Zehnder interferometer as label-free biosensor with scalable sensitivity.
Opt Lett ; 41(4): 753-6, 2016 Feb 15.
Article em En | MEDLINE | ID: mdl-26872180
The design, fabrication, and characterization of a label-free Mach-Zehnder interferometer (MZI) optical biosensor that incorporates a highly dispersive one-dimensional (1D) photonic crystal in one arm are presented. The sensitivity of this slow light MZI-based sensor scales with the length of the slow light photonic crystal region. The numerically simulated sensitivity of a MZI sensor with a 16 µm long slow light region is 115,000 rad/RIU-cm, which is sevenfold higher than traditional MZI biosensors with millimeter-length sensing regions. An experimental bulk refractive index detection sensitivity of 84,000 rad/RIU-cm is realized and nucleic acid detection is also demonstrated.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas Biossensoriais / Interferometria / Luz Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2016 Tipo de documento: Article