Your browser doesn't support javascript.
loading
Impact of and correction for instrument sensitivity drift on nanoparticle size measurements by single-particle ICP-MS.
El Hadri, Hind; Petersen, Elijah J; Winchester, Michael R.
Afiliação
  • El Hadri H; Chemical Sciences Division, Material Measurement Laboratory, National Institute of Standards and Technology (NIST), 100 Bureau Drive, Gaithersburg, MD, 20899, USA.
  • Petersen EJ; Materials Measurement Science Division, Material Measurement Laboratory, National Institute of Standards and Technology (NIST), 100 Bureau Drive, Gaithersburg, MD, 20899, USA.
  • Winchester MR; Biosystems and Biomaterials Division, Material Measurement Laboratory, National Institute of Standards and Technology (NIST), 100 Bureau Drive, Gaithersburg, MD, 20899, USA.
Anal Bioanal Chem ; 408(19): 5099-108, 2016 Jul.
Article em En | MEDLINE | ID: mdl-26894759
ABSTRACT
The effect of ICP-MS instrument sensitivity drift on the accuracy of nanoparticle (NP) size measurements using single particle (sp)ICP-MS is investigated. Theoretical modeling and experimental measurements of the impact of instrument sensitivity drift are in agreement and indicate that drift can impact the measured size of spherical NPs by up to 25 %. Given this substantial bias in the measured size, a method was developed using an internal standard to correct for the impact of drift and was shown to accurately correct for a decrease in instrument sensitivity of up to 50 % for 30 and 60 nm gold nanoparticles. Graphical Abstract Correction of nanoparticle size measurement by spICP-MS using an internal standard.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2016 Tipo de documento: Article