Your browser doesn't support javascript.
loading
MR feature tracking in patients with MRI-conditional pacing systems: The impact of pacing.
van der Graaf, A W Maurits; Bhagirath, Pranav; Scheffer, Mike G; de Medina, Ramon Robles; Götte, Marco J W.
Afiliação
  • van der Graaf AW; Department of Cardiology, Haga Teaching Hospital, The Hague, The Netherlands. a.vandergraaf@hagaziekenhuis.nl.
  • Bhagirath P; Department of Cardiology, Haga Teaching Hospital, The Hague, The Netherlands.
  • Scheffer MG; Department of Cardiology, Haga Teaching Hospital, The Hague, The Netherlands.
  • de Medina RR; Reinier de Graaf Hospital, Delft, The Netherlands.
  • Götte MJ; Department of Cardiology, Haga Teaching Hospital, The Hague, The Netherlands.
J Magn Reson Imaging ; 44(4): 964-71, 2016 10.
Article em En | MEDLINE | ID: mdl-26990922
ABSTRACT

PURPOSE:

To develop feature tracking (FT) software to perform strain analysis on conventional (nontagged) cardiac magnetic resonance imaging (MRI) function images. With the advent of MRI-conditional pacemaker systems, effects of cardiac pacing on myocardial strain can be studied using MR. In this study the impact of pacing on left ventricular (LV) strain was investigated using MR-FT in patients with an MRI-conditional cardiac implantable electronic device (CIED). MATERIALS AND

METHODS:

FT was performed on 32 1.5T MR studies (16 patients with an MRI-conditional CIED and 16 control patients with normal scans). Short- and long-axis steady state free precession (SSFP) cines were used for the FT analysis. Strain was assessed using CVI(42) software (Circle Cardiovascular Imaging, Alberta, Canada). In addition, the intra- and interobserver variability was determined using the intraclass correlation coefficient.

RESULTS:

Of the 16 patients with an MRI-conditional CIED, five patients were paced during the MRI exam. Despite the occasional presence of susceptibility artifacts induced by the CIED, radial, circumferential, and longitudinal strain parameters could be derived for all patients. Peak radial strain and peak circumferential strain were reduced during pacing when compared to the control group; for radial strain 20.1 ± 4.7% vs. 33.1 ± 6.9%, P < 0.001, and for circumferential strain -7.5 ± 3.5% vs. -14.9 ± 3.2%, P < 0.05. Peak strain parameters were reproducible on an intra- and interobserver level.

CONCLUSION:

MR-FT is feasible in patients with an MRI-conditional CIED and can be used to quantify regional wall motion. J. MAGN. RESON. IMAGING 2016;44964-971.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Assistida por Computador / Estimulação Cardíaca Artificial / Disfunção Ventricular Esquerda / Imagem Cinética por Ressonância Magnética / Técnicas de Imagem por Elasticidade Tipo de estudo: Diagnostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Assistida por Computador / Estimulação Cardíaca Artificial / Disfunção Ventricular Esquerda / Imagem Cinética por Ressonância Magnética / Técnicas de Imagem por Elasticidade Tipo de estudo: Diagnostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article