Comparison of Femoral Head Rotation and Varus Collapse Between a Single Lag Screw and Integrated Dual Screw Intertrochanteric Hip Fracture Fixation Device Using a Cadaveric Hemi-Pelvis Biomechanical Model.
Santoni, Brandon G; Nayak, Aniruddh N; Cooper, Seth A; Smithson, Ian R; Cox, Jacob L; Marberry, Scott T; Sanders, Roy W.
Afiliação
  • Santoni BG; *Foundation for Orthopaedic Research and Education, Tampa, FL;†Department of Orthopaedics and Sports Medicine, University of South Florida, Tampa, FL; and‡Orthopaedic Trauma Service, Florida Orthopaedic Institute, Tampa, FL.
J Orthop Trauma ; 30(4): 164-9, 2016 Apr.
Article em En | MEDLINE | ID: mdl-27003028
OBJECTIVE: This study compared the stabilizing effect of 2 intertrochanteric (IT) fracture fixation devices in a cadaveric hemi-pelvis biomechanical model. METHODS: Eleven pairs of cadaveric osteopenic female hemi-pelves with intact hip joint and capsular ligaments were used. An unstable IT fracture (OTA 31-A2) was created in each specimen and stabilized with a single lag screw device (Gamma 3) or an integrated dual screw (IDS) device (InterTAN). The hemi-pelves were inverted, coupled to a biaxial apparatus and subjected to 13.5 k cycles of loading (3 months) using controlled, oscillating pelvic rotation (0-90 degrees) plus cyclic axial femoral loading at a 2:1 body weight (BW) ratio. Femoral head rotation and varus collapse were monitored optoelectonically. For specimens surviving 3 months of loading, additional loading was performed in 0.25 × BW/250 cycle increments to a maximum of 4 × BW or failure. RESULTS: Femoral head rotation with IDS fixation was significantly less than the single lag screw construct after 3 months of simulated loading (P = 0.016). Maximum femoral head rotation at the end of 4 × BW loading was 7× less for the IDS construct (P = 0.006). Varus collapse was significantly less with the IDS construct over the entire loading cycle (P = 0.021). CONCLUSIONS: In this worst-case model of an osteopenic, unstable, IT fracture, the IDS construct, likely owing to its larger surface area, noncylindrical profile, and fracture compression, provided significantly greater stability and resistance to femoral head rotation and varus collapse.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Parafusos Ósseos / Fixação Interna de Fraturas / Fraturas do Quadril / Instabilidade Articular Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Aged / Female / Humans Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Parafusos Ósseos / Fixação Interna de Fraturas / Fraturas do Quadril / Instabilidade Articular Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Aged / Female / Humans Idioma: En Ano de publicação: 2016 Tipo de documento: Article