Magnetic barcode imaging for contrast agents.
Hung, Andy H; Lilley, Laura M; Hu, Fengqin; Harrison, Victoria S R; Meade, Thomas J.
Afiliação
  • Hung AH; Department of Chemistry, Molecular Biosciences, Neurobiology, Biomedical Engineering, and Radiology, Northwestern University, Evanston, Illinois, USA.
  • Lilley LM; Department of Chemistry, Molecular Biosciences, Neurobiology, Biomedical Engineering, and Radiology, Northwestern University, Evanston, Illinois, USA.
  • Hu F; College of Chemistry, Beijing Normal University, Beijing, China.
  • Harrison VS; Department of Chemistry, Molecular Biosciences, Neurobiology, Biomedical Engineering, and Radiology, Northwestern University, Evanston, Illinois, USA.
  • Meade TJ; Department of Chemistry, Molecular Biosciences, Neurobiology, Biomedical Engineering, and Radiology, Northwestern University, Evanston, Illinois, USA.
Magn Reson Med ; 77(3): 970-978, 2017 03.
Article em En | MEDLINE | ID: mdl-27062518
ABSTRACT

PURPOSE:

To demonstrate a new MR imaging approach that unambiguously identifies and quantitates contrast agents based on intrinsic agent properties such as r1 , r2 , r2*, and magnetic susceptibility. The approach is referred to as magnetic barcode imaging (MBI).

METHODS:

Targeted and bioresponsive contrast agents were imaged in agarose phantoms to generate T1 , T2 , T2*, and quantitative susceptibility maps. The parameter maps were processed by a machine learning algorithm that is trained to recognize the contrast agents based on these parameters. The output is a quantitative map of contrast agent concentration, identity, and functional state.

RESULTS:

MBI allowed the quantitative interpretation of intensities, removed confounding backgrounds, enabled contrast agent multiplexing, and unambiguously detected the activation and binding states of bioresponsive and targeted contrast agents.

CONCLUSION:

MBI has the potential to overcome significant limitations in the interpretation, quantitation, and multiplexing of contrast enhancement by MR imaging probes. Magn Reson Med 77970-978, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Imageamento por Ressonância Magnética / Espectroscopia de Ressonância Magnética / Técnicas de Sonda Molecular / Meios de Contraste / Imagem Molecular Tipo de estudo: Diagnostic_studies / Prognostic_studies Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Imageamento por Ressonância Magnética / Espectroscopia de Ressonância Magnética / Técnicas de Sonda Molecular / Meios de Contraste / Imagem Molecular Tipo de estudo: Diagnostic_studies / Prognostic_studies Idioma: En Ano de publicação: 2017 Tipo de documento: Article