Effect of Radionuclide Activity Concentration on PET-CT Image Uniformity.
Hasford, Francis; Wyk, Bronwin Van; Mabhengu, Thulani; Vangu, Mboyo Di Tamba; Kyere, Augustine Kwame; Amuasi, John Humphrey.
Afiliação
  • Hasford F; Department of Nuclear Medicine, CM Johannesburg Academic Hospital, University of the Witwatersrand, Johannesburg, South Africa; Department of Medical Physics, School of Nuclear and Allied Sciences, University of Ghana, Accra, Ghana,; Medical Radiation Physics Centre, Radiological and Medical Science
  • Wyk BV; Department of Nuclear Medicine, CM Johannesburg Academic Hospital, University of the Witwatersrand, Johannesburg, South Africa.
  • Mabhengu T; Department of Nuclear Medicine, CM Johannesburg Academic Hospital, University of the Witwatersrand, Johannesburg, South Africa.
  • Vangu MD; Department of Nuclear Medicine, CM Johannesburg Academic Hospital, University of the Witwatersrand, Johannesburg, South Africa.
  • Kyere AK; Department of Medical Physics, School of Nuclear and Allied Sciences, University of Ghana, Accra, Ghana.
  • Amuasi JH; Department of Medical Physics, School of Nuclear and Allied Sciences, University of Ghana, Accra, Ghana.
World J Nucl Med ; 15(2): 91-5, 2016.
Article em En | MEDLINE | ID: mdl-27134558
Assessment of radionuclide activity concentration on positron emission tomography-computedr tomography (PET-CT) image uniformity has been carried out quantitatively. Tomographic PET-CT images of cylindrical phantom containing F-18 fluorodeoxyglucose (FDG) activity concentration was acquired and used for the assessment. Activity concentrations were varied and PET-CT images were acquired at the constant acquisition parameters of time, matrix size, and reconstruction algorithm, respectively. Using midtransaxial image slices, quantitative index of nonuniformity (NU), and coefficient of uniformity variation were estimated for the different activity concentrations. Maximum NUs of 17.6%, 26.3%, 32.7%, 36.2%, and 38.5% were estimated for activity concentrations of 16.87 kBq/mL, 14.06 kBq/mL, 11.25 kBq/mL, 8.43 kBq/mL, and 5.62 kBq/mL, respectively. The coefficient of uniformity variation established an inverse quadratic relationship with activity concentration. Activity concentrations of 16.87 kBq/mL, 14.06 kBq/mL, 11.25 kBq/mL, 8.43 kBq/mL, and 5.62 kBq/mL produced uniformity variations of 1.47%, 2.52%, 4.23%, 5.12%, and 4.98%, respectively. Increasing activity concentration resulted in decreasing coefficient of uniformity and hence, an increase in image uniformity. The uniformity estimates compared well with the standards set internationally.