Your browser doesn't support javascript.
loading
A 16-F Sheath with Endobronchial Forceps Improves Reported Retrieval Success of Long-Dwelling "Closed Cell" Inferior Vena Cava Filter Designs.
Chick, Jeffrey Forris Beecham; Stavropoulos, S William; Shin, Benjamin J; Shlansky-Goldberg, Richard D; Mondschein, Jeffrey I; Sudheendra, Deepak; Nadolski, Gregory J; Watts, Micah M; Trerotola, Scott O.
Afiliação
  • Chick JF; Division of Interventional Radiology, Department of Radiology, Hospital of the University of Pennsylvania, Perelman School of Medicine, 1 Silverstein, 3400 Spruce Street, Philadelphia, PA 19104.
  • Stavropoulos SW; Division of Interventional Radiology, Department of Radiology, Hospital of the University of Pennsylvania, Perelman School of Medicine, 1 Silverstein, 3400 Spruce Street, Philadelphia, PA 19104.
  • Shin BJ; Division of Interventional Radiology, Department of Radiology, Hospital of the University of Pennsylvania, Perelman School of Medicine, 1 Silverstein, 3400 Spruce Street, Philadelphia, PA 19104.
  • Shlansky-Goldberg RD; Division of Interventional Radiology, Department of Radiology, Hospital of the University of Pennsylvania, Perelman School of Medicine, 1 Silverstein, 3400 Spruce Street, Philadelphia, PA 19104.
  • Mondschein JI; Division of Interventional Radiology, Department of Radiology, Hospital of the University of Pennsylvania, Perelman School of Medicine, 1 Silverstein, 3400 Spruce Street, Philadelphia, PA 19104.
  • Sudheendra D; Division of Interventional Radiology, Department of Radiology, Hospital of the University of Pennsylvania, Perelman School of Medicine, 1 Silverstein, 3400 Spruce Street, Philadelphia, PA 19104.
  • Nadolski GJ; Division of Interventional Radiology, Department of Radiology, Hospital of the University of Pennsylvania, Perelman School of Medicine, 1 Silverstein, 3400 Spruce Street, Philadelphia, PA 19104.
  • Watts MM; Division of Interventional Radiology, Department of Radiology, Hospital of the University of Pennsylvania, Perelman School of Medicine, 1 Silverstein, 3400 Spruce Street, Philadelphia, PA 19104.
  • Trerotola SO; Division of Interventional Radiology, Department of Radiology, Hospital of the University of Pennsylvania, Perelman School of Medicine, 1 Silverstein, 3400 Spruce Street, Philadelphia, PA 19104. Electronic address: scott.trerotola@uphs.upenn.edu.
J Vasc Interv Radiol ; 27(7): 1027-33, 2016 Jul.
Article em En | MEDLINE | ID: mdl-27241396
ABSTRACT

PURPOSE:

To report results of a 16-F sheath/endobronchial forceps combination in retrieval of "closed cell" inferior vena cava (IVC) filter designs with prolonged dwell times of ≥ 60 days. MATERIALS AND

METHODS:

Forceps retrieval of selected IVC filters using a 16-F 45-cm sheath was performed in 35 patients (25 women and 10 men). Patients presented with filters that were no longer needed without (n = 10) or with (n = 21) prior failed retrieval, caval thrombosis before lysis (n = 3), and back pain (n = 1). Filter designs included 19 Günther Tulip (54%), 10 Option (29%), and six OptEase (17%). Mean dwell time was 977 days. Imaging characteristics, technical success, fluoroscopy time, and complications were recorded.

RESULTS:

Filters included 20 tip-embedded, 6 tip-adherent, 4 fractured, and 6 containing thrombus. All were strut and wall-embedded. Initial snare failures occurred in 8 filters because of hook straightening (n = 4), adherent material (n = 2), snare breakage (n = 1), and filter tilt (n = 1). The 16-F sheath/forceps combination was 100% successful in removing filters. In 1 patient, 2 extravascular fragments could not be retrieved. Median fluoroscopy time, excluding 3 extended lysis procedures, was 8 minutes. Minor caval abnormalities, including caval spasm and caval defects, were noted in 17 patients, and filter fracture occurred in 2 patients. No major complications occurred.

CONCLUSIONS:

A high rate of retrieval for closed cell long-dwelling strut and wall-embedded filters may be achieved using a 16-F sheath/endobronchial forceps combination.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Instrumentos Cirúrgicos / Veia Cava Inferior / Filtros de Veia Cava / Implantação de Prótese / Remoção de Dispositivo Tipo de estudo: Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged País/Região como assunto: America do norte Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Instrumentos Cirúrgicos / Veia Cava Inferior / Filtros de Veia Cava / Implantação de Prótese / Remoção de Dispositivo Tipo de estudo: Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged País/Região como assunto: America do norte Idioma: En Ano de publicação: 2016 Tipo de documento: Article