ABSTRACT

AIMS:

Supraventricular tachycardia, mistakenly diagnosed as ventricular tachycardia, may lead to inappropriate ICD therapy. Inappropriate ICD therapy associates with increased morbidity and mortality. A reduction of inappropriate therapy is therefore desired. METHODS AND

RESULTS:

We present three different cases that beautifully illustrate the diagnostic possibilities of stored electrograms. Initial differential diagnoses in all three patients included ventricular tachycardia, focal atrial tachycardia, atrio-ventricular re-entry tachycardia, and atrio-ventricular nodal re-entry tachycardia. The first case was diagnosed as atrio-ventricular nodal re-entry tachycardia. In the second case, atrio-ventricular re-entry tachycardia was revealed. Ectopic atrial tachycardia was the reason for inappropriate therapy in the third patient.

CONCLUSION:

Thorough analysis of stored electrograms yielded correct diagnosis in all cases, which subsequently was confirmed by electrophysiological study.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Taquicardia Supraventricular / Taquicardia Ventricular / Desfibriladores Implantáveis / Erros de Diagnóstico / Eletrocardiografia Tipo de estudo: Diagnostic_studies Limite: Adult / Humans / Male / Middle aged Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Taquicardia Supraventricular / Taquicardia Ventricular / Desfibriladores Implantáveis / Erros de Diagnóstico / Eletrocardiografia Tipo de estudo: Diagnostic_studies Limite: Adult / Humans / Male / Middle aged Idioma: En Ano de publicação: 2017 Tipo de documento: Article