Feasibility of liver stiffness measurement in morbidly obese patients undergoing bariatric surgery using XL probe.
Weiss, Johannes; Rau, Monika; Meertens, Jan; Hering, Ilona; Reichert, Lisa; Kudlich, Theodor; Koschker, Ann-Kathrin; Jurowich, Christian; Geier, Andreas.
Afiliação
  • Weiss J; a Division of Hepatology , University Hospital of Wuerzburg , Germany ;
  • Rau M; a Division of Hepatology , University Hospital of Wuerzburg , Germany ;
  • Meertens J; a Division of Hepatology , University Hospital of Wuerzburg , Germany ;
  • Hering I; a Division of Hepatology , University Hospital of Wuerzburg , Germany ;
  • Reichert L; a Division of Hepatology , University Hospital of Wuerzburg , Germany ;
  • Kudlich T; b Division of Gastroenterology , University Hospital of Wuerzburg , Germany ;
  • Koschker AK; c Division of Endocrinology , University Hospital of Wuerzburg , Germany ;
  • Jurowich C; d Department of Surgery , University Hospital of Wuerzburg , Germany.
  • Geier A; a Division of Hepatology , University Hospital of Wuerzburg , Germany ;
Scand J Gastroenterol ; 51(10): 1263-8, 2016 Oct.
Article em En | MEDLINE | ID: mdl-27308959
ABSTRACT

OBJECTIVE:

Prevalence of non-alcoholic fatty liver disease is rising in the Western world and reaches up to 90% in patients undergoing bariatric surgery. Fibroscan(®) as a non-invasive tool for liver stiffness measurement (LSM) has several limitations in morbidly obese patients. Only few data exist about the technical feasibility and accuracy of LSM in these patients. We aimed to analyse the feasibility of LSM by Fibroscan(®) in bariatric patients. MATERIALS AND

METHODS:

In morbidly obese patients, LSM was performed using XL probe. Measurements were termed reliable if 10 successful measurements with a success rate ≥60% and an interquartile range/median (IQR/M) <0.3 were obtained, unreliable if 10 successful measurements were obtained but the IQR/M was >0.3, and they were termed failed if they were neither reliable nor unreliable.

RESULTS:

A total of 149 patients were included (87 with liver biopsies); mean BMI was 51.6 ± 8.5 kg/m(2). In 41% LSM using XL-probe was reliable, in 22% unreliable and in 37% failed. Failed LSM was significantly more frequent in patients with higher BMI compared to reliable and unreliable measurements (p < 0.05). In patients with failed measurement, sonographic paramedian and intercostal distances were significantly higher compared to reliable measurements. All three patients with F4 fibrosis could successfully be differentiated by LSM from patients without fibrosis.

CONCLUSIONS:

LSM with XL probe is feasible in almost two-thirds of morbidly obese patients with a BMI ≥50 kg/m(2). Reliable prediction of advanced fibrosis appears to be possible even if formal criteria of successful measurements are not met.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Obesidade Mórbida / Hepatopatia Gordurosa não Alcoólica / Fígado / Cirrose Hepática Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Obesidade Mórbida / Hepatopatia Gordurosa não Alcoólica / Fígado / Cirrose Hepática Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Ano de publicação: 2016 Tipo de documento: Article