Your browser doesn't support javascript.
loading
Accuracy and Trending of Continuous Noninvasive Hemoglobin Monitoring in Patients Undergoing Liver Transplantation.
Huang, P H; Shih, B F; Tsai, Y-F; Chung, P C H; Liu, F C; Yu, H P; Lee, W C; Chang, C J; Lin, C C.
Afiliação
  • Huang PH; Department of Anesthesiology, Chang Gung Memorial Hospital, Guishan, Taoyuan, Taiwan.
  • Shih BF; Department of Anesthesiology, Chang Gung Memorial Hospital, Guishan, Taoyuan, Taiwan.
  • Tsai YF; Department of Anesthesiology, Chang Gung Memorial Hospital, Guishan, Taoyuan, Taiwan; College of Medicine, Chang Gung University, Guishan, Taoyuan, Taiwan.
  • Chung PC; Department of Anesthesiology, Chang Gung Memorial Hospital, Guishan, Taoyuan, Taiwan; College of Medicine, Chang Gung University, Guishan, Taoyuan, Taiwan.
  • Liu FC; Department of Anesthesiology, Chang Gung Memorial Hospital, Guishan, Taoyuan, Taiwan; College of Medicine, Chang Gung University, Guishan, Taoyuan, Taiwan.
  • Yu HP; Department of Anesthesiology, Chang Gung Memorial Hospital, Guishan, Taoyuan, Taiwan; College of Medicine, Chang Gung University, Guishan, Taoyuan, Taiwan.
  • Lee WC; College of Medicine, Chang Gung University, Guishan, Taoyuan, Taiwan; Department of General Surgery, Chang Gung Memorial Hospital, Guishan, Taoyuan, Taiwan.
  • Chang CJ; Clinical Informatics and Medical Statistics Research Center, Chang Gung University, Guishan, Taoyuan, Taiwan; Biostatistical Center for Clinical Research, Chang Gung Memorial Hospital, Guishan, Taoyuan, Taiwan.
  • Lin CC; Department of Anesthesiology, Chang Gung Memorial Hospital, Guishan, Taoyuan, Taiwan; College of Medicine, Chang Gung University, Guishan, Taoyuan, Taiwan. Electronic address: chihchung@cgmh.org.tw.
Transplant Proc ; 48(4): 1067-70, 2016 May.
Article em En | MEDLINE | ID: mdl-27320558
ABSTRACT

BACKGROUND:

Shift in large fluid volumes and massive blood loss during liver transplantation frequently leads to rapid changes in hemoglobin (Hb) concentration; thus, to ensure adequate tissue oxygenation, accurate and rapid determination of Hb concentration is essential in transplant recipients. The Radical-7 Pulse CO-Oximeter provides a noninvasive and continuous way to monitor Hb concentration (SpHb) in real time and is an ideal candidate for use during liver transplantation. In this study, we assessed the relationship between SpHb and total Hb (tHb) obtained from arterial blood samples during surgery.

METHODS:

Forty patients undergoing liver transplantation were enrolled in this study. tHb and time-matched SpHb were measured at 5 different phases throughout surgery. Paired SpHb and tHb levels were assessed using linear regression, Bland-Altman analysis, and the Critchley polar plot method.

RESULTS:

A total of 161 paired measurements with sufficient signal quality were analyzed. The correlation between SpHb and tHb was 0.59 (P < .001). Bland-Altman analysis revealed that a bias between SpHb and tHb was 2.28 g/dL, and limits of agreement (LoA) were from -0.78 to 5.34 g/dL. Trending analysis showed that 87% of data were located within the acceptable trending area, indicating that the trending ability was not satisfied.

CONCLUSIONS:

The Radical-7 Pulse CO-Oximeter was not sufficient to monitor Hb levels and trends during liver transplantation surgery in our cohort. In particular, in critical patients and in those with low Hb levels, invasive Hb measurement should be used for assessment.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Hemoglobinas / Oximetria / Monitorização Intraoperatória / Transplante de Fígado Tipo de estudo: Evaluation_studies / Observational_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Hemoglobinas / Oximetria / Monitorização Intraoperatória / Transplante de Fígado Tipo de estudo: Evaluation_studies / Observational_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article