Avoiding full corrections in dynamic SPECT images impacts the performance of SPECT myocardial blood flow quantitation.
Wang, Lei; Wu, Dayong; Yang, Yong; Chen, Ing-Jou; Lin, Chih-Yuan; Hsu, Bailing; Fang, Wei; Tang, Yi-Da.
Afiliação
  • Wang L; Department of Nuclear Medicine, Fu Wai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Wu D; Department of Nuclear Medicine, Fu Wai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Yang Y; Department of Nuclear Medicine, Fu Wai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Chen IJ; Department of Medical Physics Research, Bailing Cloud Biomedical Technologies Innovation, Taipei, Taiwan.
  • Lin CY; Department of Medical Physics Research, Bailing Cloud Biomedical Technologies Innovation, Taipei, Taiwan.
  • Hsu B; Department of Electro-optical Engineering, National Taipei University of Technology, Taipei, Taiwan.
  • Fang W; Nuclear Science and Engineering Institute, University of Missouri-Columbia, E2433 Lafferre Hall, Columbia, MO, 65211, USA. bailinghsu@gmail.com.
  • Tang YD; Department of Nuclear Medicine, Fu Wai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China. nuclearfw@126.com.
J Nucl Cardiol ; 24(4): 1332-1346, 2017 08.
Article em En | MEDLINE | ID: mdl-27338944
ABSTRACT

INTRODUCTION:

This study investigated the performance of SPECT myocardial blood flow (MBF) quantitation lacking full physical corrections (All Corr) in dynamic SPECT (DySPECT) images.

METHODS:

Eleven healthy normal volunteers (HVT) and twenty-four patients with angiography-documented CAD were assessed. All Corr in 99mTc-sestamibi DySPECT encompassed noise reduction (NR), resolution recovery (RR), and corrections for scatter (SC) and attenuation (AC), otherwise no correction (NC) or only partial corrections. The performance was evaluated by quality index (R 2) and blood-pool spillover index (FBV) in kinetic modeling, and by rest flow (RMBF) and stress flow (SMBF) compared with those of All Corr.

RESULTS:

In HVT group, NC diminished 2-fold flow uniformity with the most degraded quality (15%-18% reduced R 2) and elevated spillover effect (45%-50% increased FBV). Consistently higher RMBF and SMBF were discovered in both groups (HVT 1.54/2.31 higher; CAD 1.60/1.72; all P < .0001). Bland-Altman analysis revealed positive flow bias (HVT 0.9-2.6 mL/min/g; CAD 0.7-1.3) with wide ranges of 95% CI of agreement (HVT NC -1.9-7.1; NR -0.4-4.4; NR + SC -1.1-4.3; NR + SC + RR -0.7-2.5) (CAD NC -1.2-3.8; NR -1.0-2.8; NR + SC -1.0-2.5; NR + SC + RR -1.1-2.6).

CONCLUSIONS:

Uncorrected physical interference in DySPECT images can extensively impact the performance of MBF quantitation. Full physical corrections should be considered to warrant this tool for clinical utilization.