Your browser doesn't support javascript.
loading
Development of a single-cell X-ray fluorescence flow cytometer.
Crawford, Andrew M; Kurecka, Patrick; Yim, Tsz Kwan; Kozemchak, Claire; Deb, Aniruddha; Dostál, Lubomír; Sun, Cheng Jun; Brewe, Dale L; Barrea, Raul; Penner-Hahn, James E.
Afiliação
  • Crawford AM; Department of Chemistry, University of Michigan, Ann Arbor, MI 48109-1055, USA.
  • Kurecka P; Department of Chemistry, University of Michigan, Ann Arbor, MI 48109-1055, USA.
  • Yim TK; Department of Chemistry, University of Michigan, Ann Arbor, MI 48109-1055, USA.
  • Kozemchak C; Department of Chemistry, University of Michigan, Ann Arbor, MI 48109-1055, USA.
  • Deb A; Department of Chemistry, University of Michigan, Ann Arbor, MI 48109-1055, USA.
  • Dostál L; Department of Chemistry, University of Michigan, Ann Arbor, MI 48109-1055, USA.
  • Sun CJ; X-ray Science Division, Argonne National Laboratory, 9700 South Cass Avenue, Argonne, IL 60439, USA.
  • Brewe DL; X-ray Science Division, Argonne National Laboratory, 9700 South Cass Avenue, Argonne, IL 60439, USA.
  • Barrea R; Department of Physics, DePaul University, Chicago, IL 60604, USA.
  • Penner-Hahn JE; Department of Chemistry, University of Michigan, Ann Arbor, MI 48109-1055, USA.
J Synchrotron Radiat ; 23(Pt 4): 901-8, 2016 07.
Article em En | MEDLINE | ID: mdl-27359138
ABSTRACT
An X-ray fluorescence flow cytometer that can determine the total metal content of single cells has been developed. Capillary action or pressure was used to load cells into hydrophilic or hydrophobic capillaries, respectively. Once loaded, the cells were transported at a fixed vertical velocity past a focused X-ray beam. X-ray fluorescence was then used to determine the mass of metal in each cell. By making single-cell measurements, the population heterogeneity for metals in the µM to mM concentration range on fL sample volumes can be directly measured, a measurement that is difficult using most analytical methods. This approach has been used to determine the metal composition of 936 individual bovine red blood cells (bRBC), 31 individual 3T3 mouse fibroblasts (NIH3T3) and 18 Saccharomyces cerevisiae (yeast) cells with an average measurement frequency of ∼4 cells min(-1). These data show evidence for surprisingly broad metal distributions. Details of the device design, data analysis and opportunities for further sensitivity improvement are described.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Citometria de Fluxo Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Citometria de Fluxo Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Ano de publicação: 2016 Tipo de documento: Article