A Magnetic Nanoparticle Based Nucleic Acid Isolation and Purification Instrument for DNA Extraction of Escherichia Coli O157: H7.
J Nanosci Nanotechnol
; 16(3): 2296-300, 2016 Mar.
Article
em En
| MEDLINE
| ID: mdl-27455632
The objective of this study was to evaluate the performance of a nucleic acid isolation and purification instrument using Escherichia coli O157:H7 as the model. The instrument was developed with magnetic nanoparticles for efficiently capturing nucleic acids and an intelligent mechanical unit for automatically performing the whole nucleic acid extraction process. A commercial DNA extraction kit from Huier Nano Company was used as reference. Nucleic acids in 1 ml of E. coli O157: H7 at a concentration of 5 x 10(8) CFU/mL were extracted by using this instrument and the kit in parallel and then detected by an ultraviolet spectrophotometer to obtain A260 values and A260/A280 values for the determination of the extracted DNA's quantity and purity, respectively. The A260 values for the instrument and the kit were 0.78 and 0.61, respectively, and the A260/A280 values were 1.98 and 1.93. The coefficient of variations of these parallel tests ranged from 10.5% to 16.7%. The results indicated that this nucleic acid isolation and purification instrument could extract a comparable level of nucleic acid within 50 min compared to the commercial DNA extraction kit.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
DNA Bacteriano
/
Ácidos Nucleicos
/
Escherichia coli O157
/
Nanopartículas
/
Magnetismo
Tipo de estudo:
Prognostic_studies
Idioma:
En
Ano de publicação:
2016
Tipo de documento:
Article