Improved methods for evaluating pre-clinical and histological effects of subcutaneous fat reduction using high-intensity focused ultrasound in a porcine model.
Skin Res Technol
; 23(2): 194-201, 2017 May.
Article
em En
| MEDLINE
| ID: mdl-27511869
ABSTRACT
BACKGROUND:
Non-invasive body sculpting procedures are becoming increasingly popular. High-intensity focused ultrasound (HIFU) treatment is a non-surgical fat reduction procedure that permanently destroys unwanted abdominal fat. Despite its increasing popularity, evaluation methods for the procedure have not yet been fully developed.AIMS:
The objective of this study was to develop evaluation methods for HIFU for non-surgical, permanent fat reduction in the anterior abdomen using a porcine model.METHODS:
The abdomens of female pigs (Sus scrofa, n = 7) were treated with a HIFU device (SCIZER™ , Classys Inc, Seoul, Korea). We examined treatment effects using photography, ultrasound, gross and microscopic pathology, and serum lipid and liver function level analysis, carbon tracer test, and histological examination in order to determine the mechanism of action, efficacy, and safety of the procedure.RESULTS:
HIFU treatment effectively reduced abdominal fat in a porcine model; it accurately treated the target subcutaneous fat layer and the subcutaneous fat was reduced effectively via ultrasonic measurement after HIFU treatment. On histological staining (H&E, toluidine blue, oil red O and immunohistochemistry), we found that subcutaneous fat reduction occurred effectively via accurate treatment of the targeted subcutaneous fat layer. On hematological assay, there were changes within normal range, and values remained stable after 48 h. Via carbon tracer test, the migration of activated macrophages was identified within the axillary lymph node (LN). PPAR-delta, a protein defined by immunohistochemistry staining, was overexpressed in the early stage on days 1 and 7, but a gradual decreasing pattern was confirmed.CONCLUSION:
We successfully used a HIFU device for body contouring and fat reduction in a pre-clinical study. These results provide that the essential clues toward the effective evaluation, guiding selection of the appropriate diagnostic investigations.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Lipectomia
/
Técnicas Cosméticas
/
Gordura Subcutânea
/
Ablação por Ultrassom Focalizado de Alta Intensidade
Tipo de estudo:
Diagnostic_studies
/
Prognostic_studies
Limite:
Animals
Idioma:
En
Ano de publicação:
2017
Tipo de documento:
Article