Your browser doesn't support javascript.
loading
IBIS: an OR ready open-source platform for image-guided neurosurgery.
Drouin, Simon; Kochanowska, Anna; Kersten-Oertel, Marta; Gerard, Ian J; Zelmann, Rina; De Nigris, Dante; Bériault, Silvain; Arbel, Tal; Sirhan, Denis; Sadikot, Abbas F; Hall, Jeffery A; Sinclair, David S; Petrecca, Kevin; DelMaestro, Rolando F; Collins, D Louis.
Afiliação
  • Drouin S; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada. simon.drouin@mail.mcgill.ca.
  • Kochanowska A; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • Kersten-Oertel M; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • Gerard IJ; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • Zelmann R; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • De Nigris D; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • Bériault S; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • Arbel T; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • Sirhan D; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • Sadikot AF; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • Hall JA; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • Sinclair DS; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • Petrecca K; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • DelMaestro RF; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
  • Collins DL; McConnell Brain Imaging Center, Montreal Neurological Institute and Hospital, WB 221, 3801 University Street, Montreal, QC, H3A 2B4, Canada.
Int J Comput Assist Radiol Surg ; 12(3): 363-378, 2017 Mar.
Article em En | MEDLINE | ID: mdl-27581336
ABSTRACT

PURPOSE:

Navigation systems commonly used in neurosurgery suffer from two main drawbacks (1) their accuracy degrades over the course of the operation and (2) they require the surgeon to mentally map images from the monitor to the patient. In this paper, we introduce the Intraoperative Brain Imaging System (IBIS), an open-source image-guided neurosurgery research platform that implements a novel workflow where navigation accuracy is improved using tracked intraoperative ultrasound (iUS) and the visualization of navigation information is facilitated through the use of augmented reality (AR).

METHODS:

The IBIS platform allows a surgeon to capture tracked iUS images and use them to automatically update preoperative patient models and plans through fast GPU-based reconstruction and registration methods. Navigation, resection and iUS-based brain shift correction can all be performed using an AR view. IBIS has an intuitive graphical user interface for the calibration of a US probe, a surgical pointer as well as video devices used for AR (e.g., a surgical microscope).

RESULTS:

The components of IBIS have been validated in the laboratory and evaluated in the operating room. Image-to-patient registration accuracy is on the order of [Formula see text] and can be improved with iUS to a median target registration error of 2.54 mm. The accuracy of the US probe calibration is between 0.49 and 0.82 mm. The average reprojection error of the AR system is [Formula see text]. The system has been used in the operating room for various types of surgery, including brain tumor resection, vascular neurosurgery, spine surgery and DBS electrode implantation.

CONCLUSIONS:

The IBIS platform is a validated system that allows researchers to quickly bring the results of their work into the operating room for evaluation. It is the first open-source navigation system to provide a complete solution for AR visualization.