Micro-liquid chromatography mass spectrometry for the analysis of antineoplastic drugs from wipe samples.
Anal Bioanal Chem
; 408(28): 8221-8229, 2016 Nov.
Article
em En
| MEDLINE
| ID: mdl-27655336
A fast quantification method for the determination of 11 antineoplastic drugs from wipe samples was developed using micro-scale liquid chromatography in combination with tandem mass spectrometry. The extraction efficiency from the wipes has been investigated using different extraction solvents. The results indicate that a mixture of 70/30 water/isopropanol (v/v) acidified with 0.1 % formic acid is suitable to desorb the antineoplastic drugs with sufficient recovery between 80 and 120 %. Compared to conventional liquid chromatography, the total analysis time can be reduced to 2.25 min using a 50 × 0.3 mm column at a flow rate of 25 µL min-1. Ion source parameters as well as the injection volume were optimized to ensure the highest sensitivity. The results of method validation showed an instrumental limit of quantification between 0.0068 and 0.0488 ng mL-1 using an injection volume of 4.25 µL estimated by the signal to noise ratio. Moreover, the retention time repeatability was determined with a maximum relative standard deviation of 0.4 %. Graphical abstract Micro-LC-MS/MS separation of 11 antineoplastic drugs from wipe samples.
Palavras-chave
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Monitoramento Ambiental
/
Cromatografia Líquida
/
Técnicas Analíticas Microfluídicas
/
Poluentes Ambientais
/
Espectrometria de Massas em Tandem
/
Antineoplásicos
Idioma:
En
Ano de publicação:
2016
Tipo de documento:
Article