Level of agreement between cardiac output measurements using Nexfin® and thermodilution in morbidly obese patients undergoing laparoscopic surgery.
Anaesthesia
; 71(12): 1449-1455, 2016 12.
Article
em En
| MEDLINE
| ID: mdl-27667644
Morbidly obese patients are at increased risk of intra-operative haemodynamic instability, which may necessitate intensive monitoring. Non-invasive monitoring is increasingly used to measure cardiac output; however, it is unknown whether the weight-based algorithm utilised in these devices is applicable to patients with morbid obesity. We compared the level of agreement and trending ability of non-invasive cardiac output measurements (Nexfin® ) with the gold-standard thermodilution technique in 30 morbidly obese patients undergoing laparoscopic surgery. Bland-Altman analysis revealed a mean (SD) bias of 0.60 (1.62) l.min-1 (limits of agreement -2.67 to 3.86 l.min-1 ) and the precision error was 46%. Polar plot analysis resulted in an angular bias of 2.61°, radial limits of agreement of -60.08° to 49.82° and angular concordance rate was 77%. Both agreement and trending were outside the Critchley criteria for the comparison of cardiac output devices with a gold-standard. Nexfin has an unacceptable level of agreement compared with thermodilution for cardiac output measurement in morbidly obese patients.