Clinical Results and Mechanical Properties of the Carotid CGUARD Double-Layered Embolic Prevention Stent.
Wissgott, Christian; Schmidt, Wolfram; Brandt-Wunderlich, Christoph; Behrens, Peter; Andresen, Reimer.
Afiliação
  • Wissgott C; 1 Institute of Diagnostic and Interventional Radiology/Neuroradiology, Westkuestenklinikum Heide-Academic Teaching Hospital of the Universities of Kiel, Luebeck, and Hamburg, Heide, Germany.
  • Schmidt W; 2 Institute for Biomedical Engineering, Rostock University Medical Center, Rostock, Germany.
  • Brandt-Wunderlich C; 2 Institute for Biomedical Engineering, Rostock University Medical Center, Rostock, Germany.
  • Behrens P; 2 Institute for Biomedical Engineering, Rostock University Medical Center, Rostock, Germany.
  • Andresen R; 1 Institute of Diagnostic and Interventional Radiology/Neuroradiology, Westkuestenklinikum Heide-Academic Teaching Hospital of the Universities of Kiel, Luebeck, and Hamburg, Heide, Germany.
J Endovasc Ther ; 24(1): 130-137, 2017 02.
Article em En | MEDLINE | ID: mdl-27733689
ABSTRACT

PURPOSE:

To report early clinical outcomes with a novel double-layer stent for the internal carotid artery (ICA) and the in vitro investigation of the stent's mechanical properties.

METHODS:

A prospective single-center study enrolled 30 consecutive patients (mean age 73.1±6.3 years; 21 men) with symptomatic (n=25) or high-grade (n=5) ICA stenosis treated with the new double-layer carotid CGUARD Embolic Prevention System (EPS) stent, which has an inner open-cell nitinol design with an outer closed-cell polyethylene terephthalate layer. The average stenosis of the treated arteries was 84.1%±7.9% with a mean lesion length of 16.6±2.1 mm. In the laboratory, 8×40-mm stents where tested in vitro with respect to their radial force during expansion, the bending stiffness of the stent system and the expanded stent, as well as the collapse pressure in a thin and flexible sheath. The wall adaptation was assessed using fluoroscopy after stent release in step and curved vessel models.

RESULTS:

The stent was successfully implanted in all patients. No peri- or postprocedural complications occurred; no minor or major stroke was observed in the 6-month follow-up. The bending stiffness of the expanded stent was 63.1 N·mm2 and (not unexpectedly) was clearly lower than that of the stent system (601.5 N·mm2). The normalized radial force during expansion of the stent to 7.0 mm, consistent with in vivo sizing, was relatively high (0.056 N/mm), which correlates well with the collapse pressure of 0.17 bars. Vessel wall adaptation was harmonic and caused no straightening of the vessel after clinical application.

CONCLUSION:

Because of its structure, the novel CGUARD EPS stent is characterized by a high flexibility combined with a high radial force and very good plaque coverage. These first clinical results demonstrate a very safe implantation behavior without any stroke up to 6 months after the procedure.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artéria Carótida Interna / Stents / Estenose das Carótidas / Angioplastia com Balão / Dispositivos de Proteção Embólica Tipo de estudo: Observational_studies / Prognostic_studies Limite: Aged / Female / Humans / Male País/Região como assunto: Europa Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artéria Carótida Interna / Stents / Estenose das Carótidas / Angioplastia com Balão / Dispositivos de Proteção Embólica Tipo de estudo: Observational_studies / Prognostic_studies Limite: Aged / Female / Humans / Male País/Região como assunto: Europa Idioma: En Ano de publicação: 2017 Tipo de documento: Article