Clinical hemodynamic evaluation of patients implanted with a fully magnetically levitated left ventricular assist device (HeartMate 3).
Uriel, Nir; Adatya, Sirtaz; Malý, Jirí; Kruse, Eric; Rodgers, Daniel; Heatley, Gerald; Herman, Ales; Sood, Poornima; Berliner, Dominik; Bauersachs, Johann; Haverich, Axel; Zelízko, Michael; Schmitto, Jan D; Netuka, Ivan.
Afiliação
  • Uriel N; Department of Medicine, University of Chicago Medical Center, Chicago, Illinois, USA. Electronic address: nuriel@medicine.bsd.uchicago.edu.
  • Adatya S; Department of Medicine, University of Chicago Medical Center, Chicago, Illinois, USA.
  • Malý J; Department of Cardiovascular and Transplant Surgery, Institute for Clinical and Experimental Medicine, Prague, Czech Republic.
  • Kruse E; Department of Medicine, University of Chicago Medical Center, Chicago, Illinois, USA.
  • Rodgers D; Department of Medicine, University of Chicago Medical Center, Chicago, Illinois, USA.
  • Heatley G; St. Jude Medical, Minneapolis, Minnesota, USA.
  • Herman A; Department of Cardiovascular and Transplant Surgery, Institute for Clinical and Experimental Medicine, Prague, Czech Republic.
  • Sood P; St. Jude Medical, Minneapolis, Minnesota, USA.
  • Berliner D; Department of Cardiothoracic, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Bauersachs J; Department of Cardiothoracic, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Haverich A; Department of Cardiothoracic, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Zelízko M; Department of Cardiovascular and Transplant Surgery, Institute for Clinical and Experimental Medicine, Prague, Czech Republic.
  • Schmitto JD; Department of Cardiothoracic, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Netuka I; Department of Cardiovascular and Transplant Surgery, Institute for Clinical and Experimental Medicine, Prague, Czech Republic.
J Heart Lung Transplant ; 36(1): 28-35, 2017 01.
Article em En | MEDLINE | ID: mdl-27773456
ABSTRACT

BACKGROUND:

The HeartMate 3 (HM3) is a Conformiteé Européenne (CE) mark-approved left ventricular assist device (LVAD) with a fully magnetically levitated rotor with features consisting of a wide range of operational speeds, wide flow paths and an artificial pulse. We performed a hemodynamic and echocardiographic evaluation of patients implanted with the HM3 LVAD to assess the speed range for optimal hemodynamic support.

METHODS:

Sixteen HM3 patients underwent pump speed ramp tests with right heart catheterization (including central venous pressure [CVP], pulmonary artery pressure, pulmonary capillary wedge pressure [PCWP] and blood pressure [BP]) and 3-dimensional echocardiography (3DE). Data were recorded at up to 13 speed settings. Speed changes were in steps of 100 revolutions per minute (rpm), starting at 4,600 rpm and ramping up to 6,200 rpm.

RESULTS:

Mean original speed was 5,306 ± 148 rpm, with a majority of patients (10 of 16, 62.5%) having normal CVPs and PCWPs at their original rpm settings. Going from lowest to highest speeds, cardiac output improved at the rate of 0.08 ± 0.08 liter/min per 100 rpm (total change 1.25 ± 1.20 liters/min) and PCWP decreased at the rate of -0.48 ± 0.27 mm Hg per 100 rpm (total change -6.13 ± 3.72 mm Hg). CVP and systolic BP did not change significantly with changes in rpm. Left ventricular end-diastolic dimension (LVEDD) decreased at a rate of -0.15 ± 0.09 cm per 100 rpm. Number of rpm was adjusted based on test results to achieve CVPs and PCWPs as close to normal limits as possible, which was feasible in 13 (81.3%) patients. For the remaining 3 patients, medical management was pursued to optimize hemodynamic support.

CONCLUSION:

Hemodynamic normalization of pressures was achieved in the majority of patients implanted with the HM3 pump within a narrow speed range.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Coração Auxiliar / Função Ventricular Esquerda / Insuficiência Cardíaca / Hemodinâmica / Magnetismo Tipo de estudo: Clinical_trials / Observational_studies / Prognostic_studies Limite: Aged / Female / Humans / Male Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Coração Auxiliar / Função Ventricular Esquerda / Insuficiência Cardíaca / Hemodinâmica / Magnetismo Tipo de estudo: Clinical_trials / Observational_studies / Prognostic_studies Limite: Aged / Female / Humans / Male Idioma: En Ano de publicação: 2017 Tipo de documento: Article