Determination of Trace Cr(VI) with Diphenylcarbazide by µFIA-Thermal Lens Microscopy.
Acta Chim Slov ; 63(4): 772-780, 2016 12.
Article em En | MEDLINE | ID: mdl-28004088
The optimum reaction parameters for the interaction of hexavalent chromium [Cr(VI)] with diphenylcarbazide in microfluidic chips (µFIA) with thermal-lens microscopic detection were selected. The characteristic feature of the applied flow scheme is the injection of the reagent into the stream containing the test metal, which enables in-field and real-time monitoring of Cr(VI) simply by flowing the sample continuously through the microchip. The limit of detection of Cr(VI) under the selected conditions (signal generating wavelength, 514.5 nm; excitation power, 100 mW; detection position, 10 cm downstream from the mixing zone of the microchip; flow rate 10 µL min-1; injection volume, 1.4 µL) is 15 ng mL-1 (2.9 × 10-7 mol L-1). The linear range is 40 ng mL-1 - 10 µg mL-1 with a relative standard deviation no higher than 10% in the concentration range 0.1-1 µg mL-1. The online monitoring by this scheme provides the possibility of up to 360 analyses per hour.
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2016 Tipo de documento: Article