Your browser doesn't support javascript.
loading
Thermal Variation During Osteotomy With Different Dental Implant Drills: A Standardized Study in Bovine Ribs.
Hochscheidt, Celso João; Shimizu, Roberto Hideo; Andrighetto, Augusto Ricardo; Moura, Luís Mauro; Golin, Alexsander Luiz; Hochscheidt, Regina Célia.
Afiliação
  • Hochscheidt CJ; *Master of Sciences in Implant Dentistry, Instituto Latino Americano de Pesquisa e Ensino Odontológico (ILAPEO), Curitiba, PR, Brazil; Technical Manager, Clínica Top Odontologia, Ponta Grossa, PR, Brazil. †Doctor of Sciences in Orthodontics, Universidade Estadual Paulista (UNESP), Araraquara, SP, Brazil; Coordinator, Master's Program in Orthodontics; Professor, ILAPEO, Curitiba, PR, Brazil. ‡Doctor of Sciences in Orthodontics, Universidade de São Paulo (USP), São Paulo, SP, Brazil; Professor, Gr
Implant Dent ; 26(1): 73-79, 2017 Feb.
Article em En | MEDLINE | ID: mdl-28114241
ABSTRACT

INTRODUCTION:

This study evaluated thermal variation (heat generation) by dental drills of similar geometries, made from different materials, during performance of 450 standardized osteotomies in bovine ribs to simulate implant site preparations. MATERIALS AND

METHODS:

Each of the 3 groups-steel with a diamond-like carbon coating (SG), experimentally surface-treated steel (EG), and aluminum-toughened zirconium ceramic (ZG)-included 3 drills, distributed across 6 subgroups. An implant motor provided torque, rotation, and irrigation for 50% of subgroups.

RESULTS:

In preparations drilled to 5 mm under irrigation, the number of osteotomies, material, and diameter did not influence thermal variation (ΔT); when drilling to 13 mm, the greatest ΔT occurred in EG (2.8°C). When drilling to 5 mm without irrigation, drill material influenced mean temperature; when drilling to 13 mm without irrigation, the greatest ΔT (5.5°C) occurred in EG. The highest temperature was observed in SG (79.6°C).

CONCLUSION:

After 50 uses, the greatest ΔT occurred in EG and the ∅3.0-mm drill exhibited the highest mean and peak temperatures. Temperatures were influenced by irrigation, number of uses, depth, drill diameter, geometry, and material; however, mean values were within physiological limits.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Osteotomia / Costelas / Implantação Dentária Endóssea / Instrumentos Odontológicos Limite: Animals Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Osteotomia / Costelas / Implantação Dentária Endóssea / Instrumentos Odontológicos Limite: Animals Idioma: En Ano de publicação: 2017 Tipo de documento: Article