Your browser doesn't support javascript.
loading
Evaluation of the new C-arm guiding system ClearGuide® in an orthopaedic and trauma operating theatre.
Müller, Marcus Christian; Frege, Sophie; Strauss, Andreas Christian; Gathen, Martin; Windemuth, Michael; Striepens, Eva Nadine.
Afiliação
  • Müller MC; Department of Orthopaedics and Trauma Surgery, University Hospital of Bonn, Sigmund-Freud-Strasse 25, 53105, Bonn, Germany.
  • Frege S; Department of Orthopaedics and Trauma Surgery, University Hospital of Bonn, Sigmund-Freud-Strasse 25, 53105, Bonn, Germany.
  • Strauss AC; Department of Orthopaedics and Trauma Surgery, University Hospital of Bonn, Sigmund-Freud-Strasse 25, 53105, Bonn, Germany.
  • Gathen M; Department of Orthopaedics and Trauma Surgery, University Hospital of Bonn, Sigmund-Freud-Strasse 25, 53105, Bonn, Germany.
  • Windemuth M; Department of Orthopaedics and Trauma Surgery, University Hospital of Bonn, Sigmund-Freud-Strasse 25, 53105, Bonn, Germany.
  • Striepens EN; Department of Psychiatry and Psychotherapy, University Hospital of Bonn, Sigmund-Freud-Strasse 25, 53105, Bonn, Germany.
Int J Med Robot ; 13(3)2017 Sep.
Article em En | MEDLINE | ID: mdl-28371198
ABSTRACT

BACKGROUND:

The objective was to evaluate whether the new intraoperative C-arm guiding system ClearGuide® (CG) reduces radiation exposure of the staff in an Orthopaedic and Trauma operation theatre.

METHODS:

Data of 95 patients CG was used were retrospectively compared using matched-pair analysis with controls without CG. Radiation dose (RD), fluoroscopic time (FT) and operation time (OT) were analysed in ten types of operative procedures.

RESULTS:

Use of CG led to a significant reduction (p ≤ 0.05) of the RD in intramedullary nailing and plate fixation of femoral shaft fractures as well as plating of tibia shaft fractures. Concerning FT, use of CG led to a significant reduction (p ≤ 0.05) while performing kyphoplasties and plate fixation of femoral shaft fractures. Regarding OT, no statistical significance was observed.

CONCLUSIONS:

CG as a simple, reproducible and intuitive communication tool for C-arm guidance reduces intraoperative staff radiation exposure especially while fixation of long bone fractures and in spine surgery.