[Comparison of 2 immunoenzyme methods for the determination of anti-cytomegalovirus IgM]. / Comparaison de deux méthodes immunoenzymatiques pour la détermination des IgM anti-cytomegalovirus.
Ann Biol Clin (Paris)
; 46(2): 119-22, 1988.
Article
em Fr
| MEDLINE
| ID: mdl-2837928
ABSTRACT
An antibody capture assay technique using a labeled antigen (Wellcome Laboratories) and an ELISA technique performed after elimination of serum IgG (Behring Laboratories) were assessed for determination of anti-cytomegalovirus IgM. Their specificity and sensitivity were compared in tests on 208 sera, some (60) of which had IgM rheumatoid factor, antinuclear antibodies or anti-Epstein-Barr virus antibodies. The other sera were analyzed relative to diagnosis of a cytomegalovirus infection. The antibody capture assay technique was slightly more sensitive than the ELISA technique but did not resolve all the specificity problems.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Imunoglobulina M
/
Ensaio de Imunoadsorção Enzimática
/
Técnicas Imunoenzimáticas
/
Citomegalovirus
/
Anticorpos Antivirais
Tipo de estudo:
Diagnostic_studies
/
Evaluation_studies
Limite:
Humans
Idioma:
Fr
Ano de publicação:
1988
Tipo de documento:
Article