Effects of implantable cardioverter/defibrillator shock and antitachycardia pacing on anxiety and quality of life: A MADIT-RIT substudy.
Perini, Alessandro Paoletti; Kutyifa, Valentina; Veazie, Peter; Daubert, James P; Schuger, Claudio; Zareba, Wojciech; McNitt, Scott; Rosero, Spencer; Tompkins, Christine; Padeletti, Luigi; Moss, Arthur J.
Afiliação
  • Perini AP; University of Florence, Florence, Italy.
  • Kutyifa V; University of Rochester Medical Center, Heart Research Follow-Up Program, Rochester, NY. Electronic address: Valentina.Kutyifa@heart.rochester.edu.
  • Veazie P; University of Rochester Medical Center, Heart Research Follow-Up Program, Rochester, NY.
  • Daubert JP; Duke University Medical Center, Durham, NC.
  • Schuger C; The Henry Ford Hospital, Cardiac Electrophysiology, Detroit, MI.
  • Zareba W; University of Rochester Medical Center, Heart Research Follow-Up Program, Rochester, NY.
  • McNitt S; University of Rochester Medical Center, Heart Research Follow-Up Program, Rochester, NY.
  • Rosero S; University of Rochester Medical Center, Heart Research Follow-Up Program, Rochester, NY.
  • Tompkins C; University of Rochester Medical Center, Heart Research Follow-Up Program, Rochester, NY.
  • Padeletti L; University of Florence, Florence, Italy; IRCCS Multimedica, Sesto San Giovanni, Milan, Italy.
  • Moss AJ; University of Rochester Medical Center, Heart Research Follow-Up Program, Rochester, NY.
Am Heart J ; 189: 75-84, 2017 Jul.
Article em En | MEDLINE | ID: mdl-28625384
ABSTRACT
Effects of implantable cardioverter/defibrillator (ICD) shocks and antitachycardia pacing (ATP) on anxiety and quality of life (QoL) in ICD patients are poorly understood.

METHODS:

We evaluated changes in QoL from baseline to 9-month follow-up using the EQ-5D questionnaire in patients enrolled in the Multicenter Automatic Defibrillator Implantation Trial-Reduce Inappropriate Therapy (MADIT-RIT) (n=1,268). We assessed anxiety levels using the Florida Shock Anxiety Scale (1-10 score) in patients with appropriate or inappropriate shocks or ATP compared to those with no ICD therapy during the first 9 months postimplant. The analysis was stratified by number of ATP or shocks (0-1 vs ≥2) and adjusted for covariates.

RESULTS:

In MADIT-RIT, 15 patients (1%) had ≥2 appropriate shocks, 38 (3%) had ≥2 appropriate ATPs. Two or more inappropriate shocks were delivered in 16 patients (1%); ≥2 inappropriate ATPs, in 70. In multivariable analysis, patients with ≥2 appropriate shocks had higher levels of shock-related anxiety than those with ≤1 appropriate shock (P<.01). Furthermore, ≥2 inappropriate shocks produced more anxiety than ≤1 inappropriate shock (P=.005). Consistently, ≥2 appropriate ATPs resulted in more anxiety than ≤1 (P=.028), whereas the number of inappropriate ATPs showed no association with anxiety levels (P=.997). However, there was no association between QoL and appropriate or inappropriate ATP/shock (all P values > .05).

CONCLUSIONS:

In MADIT-RIT, ≥2 appropriate or inappropriate ICD shocks and ≥2 appropriate ATPs are associated with more anxiety at 9-month follow-up despite no significant changes in the assessment of global QoL by the EQ-5D questionnaire. Innovative ICD programming reducing inappropriate therapies may help deal with patient concerns about the device.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ansiedade / Qualidade de Vida / Taquicardia Ventricular / Desfibriladores Implantáveis Tipo de estudo: Clinical_trials / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ansiedade / Qualidade de Vida / Taquicardia Ventricular / Desfibriladores Implantáveis Tipo de estudo: Clinical_trials / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2017 Tipo de documento: Article