Misfit and fracture load of implant-supported monolithic crowns in zirconia-reinforced lithium silicate.
Gomes, Rafael Soares; Souza, Caroline Mathias Carvalho de; Bergamo, Edmara Tatiely Pedroso; Bordin, Dimorvan; Del Bel Cury, Altair Antoninha.
Afiliação
  • Gomes RS; Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba, Departamento de Prótese e Periodontia, Piracicaba, SP, Brasil.
  • Souza CMC; Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba, Departamento de Odontologia Restauradora, Piracicaba, SP, Brasil.
  • Bergamo ETP; Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba, Departamento de Prótese e Periodontia, Piracicaba, SP, Brasil.
  • Bordin D; Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba, Departamento de Prótese e Periodontia, Piracicaba, SP, Brasil.
  • Del Bel Cury AA; Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba, Departamento de Prótese e Periodontia, Piracicaba, SP, Brasil.
J Appl Oral Sci ; 25(3): 282-289, 2017.
Article em En | MEDLINE | ID: mdl-28678947
ABSTRACT

OBJECTIVE:

In this study, marginal and internal misfit and fracture load with and without thermal-mechanical aging (TMA) of monolithic ZLS and lithium disilicate (LDS) crowns were evaluated. MATERIAL AND

METHODS:

Crowns were milled using a computer-aided design/computer-aided manufacturing system. Marginal gaps (MGs), absolute marginal discrepancy (AMD), axial gaps, and occlusal gaps were measured by X-ray microtomography (n=8). For fracture load testing, crowns were cemented in a universal abutment, and divided into four groups ZLS without TMA, ZLS with TMA, LDS without TMA, and LDS with TMA (n=10). TMA groups were subjected to 10,000 thermal cycles (5-55°C) and 1,000,000 mechanical cycles (200 N, 3.8 Hz). All groups were subjected to compressive strength testing in a universal testing machine at a crosshead speed of 1 mm/min until failure. Student's t-test was used to examine misfit, two-way analysis of variance was used to analyze fracture load, and Pearson's correlation coefficients for misfit and fracture load were calculated (α=0.05). The materials were analyzed according to Weibull distribution, with 95% confidence intervals.

RESULTS:

Average MG (p<0.001) and AMD (p=0.003) values were greater in ZLS than in LDS crowns. TMA did not affect the fracture load of either material. However, fracture loads of ZLS crowns were lower than those of LDS crowns (p<0.001). Fracture load was moderately correlated with MG (r=-0.553) and AMD (r=-0.497). ZLS with TMA was least reliable, according to Weibull probability.

CONCLUSION:

Within the limitations of this study, ZLS crowns had lower fracture load values and greater marginal misfit than did LDS crowns, although these values were within acceptable limits.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Compostos Organometálicos / Succinatos / Zircônio / Adaptação Marginal Dentária / Prótese Dentária Fixada por Implante / Falha de Restauração Dentária / Coroas Tipo de estudo: Evaluation_studies Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Compostos Organometálicos / Succinatos / Zircônio / Adaptação Marginal Dentária / Prótese Dentária Fixada por Implante / Falha de Restauração Dentária / Coroas Tipo de estudo: Evaluation_studies Idioma: En Ano de publicação: 2017 Tipo de documento: Article