Your browser doesn't support javascript.
loading
Automated IMRT planning in Pinnacle : A study in head-and-neck cancer.
Kusters, J M A M; Bzdusek, K; Kumar, P; van Kollenburg, P G M; Kunze-Busch, M C; Wendling, M; Dijkema, T; Kaanders, J H A M.
Afiliação
  • Kusters JMAM; Department of Radiation Oncology, Radboud university medical center, Postbox 9101, 6500 HB, Nijmegen, The Netherlands. martijn.kusters@radboudumc.nl.
  • Bzdusek K; Philips Radiation Oncology Systems, Philips Healthcare, 53711, Fitchburg, WI, USA.
  • Kumar P; Philips Innovation Campus, Philips Electronics India Ltd., Bangalore, India.
  • van Kollenburg PGM; Department of Radiation Oncology, Radboud university medical center, Postbox 9101, 6500 HB, Nijmegen, The Netherlands.
  • Kunze-Busch MC; Department of Radiation Oncology, Radboud university medical center, Postbox 9101, 6500 HB, Nijmegen, The Netherlands.
  • Wendling M; Department of Radiation Oncology, Radboud university medical center, Postbox 9101, 6500 HB, Nijmegen, The Netherlands.
  • Dijkema T; Department of Radiation Oncology, Radboud university medical center, Postbox 9101, 6500 HB, Nijmegen, The Netherlands.
  • Kaanders JHAM; Department of Radiation Oncology, Radboud university medical center, Postbox 9101, 6500 HB, Nijmegen, The Netherlands.
Strahlenther Onkol ; 193(12): 1031-1038, 2017 Dec.
Article em En | MEDLINE | ID: mdl-28770294
ABSTRACT

PURPOSE:

This study evaluates the performance and planning efficacy of the Auto-Planning (AP) module in the clinical version of Pinnacle 9.10 (Philips Radiation Oncology Systems, Fitchburg, WI, USA). METHODS AND MATERIALS Twenty automated intensity-modulated radiotherapy (IMRT) plans were compared with the original manually planned clinical IMRT plans from patients with oropharyngeal cancer.

RESULTS:

Auto-Planning with IMRT offers similar coverage of the planning target volume as the original manually planned clinical plans, as well as better sparing of the contralateral parotid gland, contralateral submandibular gland, larynx, mandible, and brainstem. The mean dose of the contralateral parotid gland and contralateral submandibular gland could be reduced by 2.5 Gy and 1.7 Gy on average. The number of monitor units was reduced with an average of 143.9 (18%). Hands-on planning time was reduced from 1.5-3 h to less than 1 h.

CONCLUSIONS:

The Auto-Planning module was able to produce clinically acceptable head and neck IMRT plans with consistent quality.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Software / Neoplasias Orofaríngeas / Radioterapia Conformacional / Órgãos em Risco Tipo de estudo: Diagnostic_studies / Etiology_studies / Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Software / Neoplasias Orofaríngeas / Radioterapia Conformacional / Órgãos em Risco Tipo de estudo: Diagnostic_studies / Etiology_studies / Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2017 Tipo de documento: Article