Your browser doesn't support javascript.
loading
Hemodynamic Evaluation of a Biological and Mechanical Aortic Valve Prosthesis Using Patient-Specific MRI-Based CFD.
Hellmeier, Florian; Nordmeyer, Sarah; Yevtushenko, Pavlo; Bruening, Jan; Berger, Felix; Kuehne, Titus; Goubergrits, Leonid; Kelm, Marcus.
Afiliação
  • Hellmeier F; Biofluid Mechanics Laboratory, Charité - Universitätsmedizin Berlin, Berlin, Germany.
  • Nordmeyer S; Institute for Computational and Imaging Science in Cardiovascular Medicine, Charité - Universitätsmedizin Berlin, Berlin, Germany.
  • Yevtushenko P; Department of Congenital Heart Disease and Pediatric Cardiology, Deutsches Herzzentrum Berlin, Berlin, Germany.
  • Bruening J; Biofluid Mechanics Laboratory, Charité - Universitätsmedizin Berlin, Berlin, Germany.
  • Berger F; Biofluid Mechanics Laboratory, Charité - Universitätsmedizin Berlin, Berlin, Germany.
  • Kuehne T; Institute for Computational and Imaging Science in Cardiovascular Medicine, Charité - Universitätsmedizin Berlin, Berlin, Germany.
  • Goubergrits L; Department of Congenital Heart Disease and Pediatric Cardiology, Deutsches Herzzentrum Berlin, Berlin, Germany.
  • Kelm M; Institute for Computational and Imaging Science in Cardiovascular Medicine, Charité - Universitätsmedizin Berlin, Berlin, Germany.
Artif Organs ; 42(1): 49-57, 2018 Jan.
Article em En | MEDLINE | ID: mdl-28853220
ABSTRACT
Modeling different treatment options before a procedure is performed is a promising approach for surgical decision making and patient care in heart valve disease. This study investigated the hemodynamic impact of different prostheses through patient-specific MRI-based CFD simulations. Ten time-resolved MRI data sets with and without velocity encoding were obtained to reconstruct the aorta and set hemodynamic boundary conditions for simulations. Aortic hemodynamics after virtual valve replacement with a biological and mechanical valve prosthesis were investigated. Wall shear stress (WSS), secondary flow degree (SFD), transvalvular pressure drop (TPD), turbulent kinetic energy (TKE), and normalized flow displacement (NFD) were evaluated to characterize valve-induced hemodynamics. The biological prostheses induced significantly higher WSS (medians 9.3 vs. 8.6 Pa, P = 0.027) and SFD (means 0.78 vs. 0.49, P = 0.002) in the ascending aorta, TPD (medians 11.4 vs. 2.7 mm Hg, P = 0.002), TKE (means 400 vs. 283 cm2 /s2 , P = 0.037), and NFD (means 0.0994 vs. 0.0607, P = 0.020) than the mechanical prostheses. The differences between the prosthesis types showed great inter-patient variability, however. Given this variability, a patient-specific evaluation is warranted. In conclusion, MRI-based CFD offers an opportunity to assess the interactions between prosthesis and patient-specific boundary conditions, which may help in optimizing surgical decision making and providing additional guidance to clinicians.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Desenho de Prótese / Implante de Prótese de Valva Cardíaca / Doenças das Valvas Cardíacas / Modelos Cardiovasculares Tipo de estudo: Guideline / Prognostic_studies Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Desenho de Prótese / Implante de Prótese de Valva Cardíaca / Doenças das Valvas Cardíacas / Modelos Cardiovasculares Tipo de estudo: Guideline / Prognostic_studies Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article