Clinical Tremor Severity Estimation Using an Instrumented Eating Utensil.
J Parkinsons Dis
; 7(4): 755-759, 2017.
Article
em En
| MEDLINE
| ID: mdl-28922166
ABSTRACT
We demonstrate the feasibility of estimating clinical tremor scores using an eating utensil with motion-sensing and tremor-cancellation technology in thirteen patients with tremor. Three experts scored hand tremor using the modified Fahn- Tolosa-Marin (FTM) scale. A linear model was trained to estimate tremor severity using the recorded motion signals. The average neurologist FTM score was 1.6±0.7 for PD and 2.6±0.7 for ET patients. The average model score was 1.6±0.7 for PD and 2.6±0.6 for ET. Correlation coefficient between the clinical and model tremor scores was 0.91 (pâ<â0.001). Motion data from an instrumented eating utensil accurately derived tremor ratings enabling practical, objective daily monitoring.