Computation of Pacemakers Immunity to 50 Hz Electric Field: Induced Voltages 10 Times Greater in Unipolar Than in Bipolar Detection Mode.
Gercek, Cihan; Kourtiche, Djilali; Nadi, Mustapha; Magne, Isabelle; Schmitt, Pierre; Souques, Martine.
Afiliação
  • Gercek C; Institut Jean Lamour, Université de Lorraine-CNRS (UMR 7198), BP 70239, 54506 Vandœuvre, France. cihan.gercek@univ-lorraine.fr.
  • Kourtiche D; Institut Jean Lamour, Université de Lorraine-CNRS (UMR 7198), BP 70239, 54506 Vandœuvre, France. djilali.kourtiche@univ-lorraine.fr.
  • Nadi M; Institut Jean Lamour, Université de Lorraine-CNRS (UMR 7198), BP 70239, 54506 Vandœuvre, France. mustapha.nadi@univ-lorraine.fr.
  • Magne I; EDF R&D, Avenue des Renardières-Ecuelles, 77818 Moret-sur-Loing, France. isabelle.magne@edf.fr.
  • Schmitt P; Institut Jean Lamour, Université de Lorraine-CNRS (UMR 7198), BP 70239, 54506 Vandœuvre, France. pierre.schmitt@univ-lorraine.fr.
  • Souques M; EDF SEM, Immeuble Carré Vert, 45 rue Kleber, 92309 Levallois-Perret CEDEX, France. martine.souques@edf.fr.
Bioengineering (Basel) ; 4(1)2017 Mar 06.
Article em En | MEDLINE | ID: mdl-28952498
ABSTRACT
Thisstudy aims to compute 50 Hz electric field interferences on pacemakers for diverse lead configurations and implantation positions. Induced phenomena in a surface-based virtual human model (standing male grounded with arms closed, 2 mm resolution) are computed for vertical exposure using CST EM® 3D software, with and without an implanted pacemaker. Induced interference voltages occurring on the pacemaker during exposure are computed and the results are discussed. The bipolar mode covers 99% of the implanted pacing leads in the USA and Europe, according to statistics. The tip-to-ring distance of a lead may influence up to 46% of the induced voltage. In bipolar sensing mode, right ventricle implantation has a 41% higher induced voltage than right atrium implantation. The induced voltage is in average 10 times greater in unipolar mode than in bipolar mode, when implanted in the right atrium or right ventricle. The electric field threshold of interference for a bipolar sensing mode in the worst case setting is 7.24 kV·m-1, and 10 times higher for nominal settings. These calculations will be completed by an in vitro study.