Influence of the Pixel Sizes of Reference Computed Tomography on Single-photon Emission Computed Tomography Image Reconstruction Using Conjugate-gradient Algorithm.
Okuda, Kyohei; Sakimoto, Shota; Fujii, Susumu; Ida, Tomonobu; Moriyama, Shigeru.
Afiliação
  • Okuda K; Division of Clinical Radiology, Tottori University Hospital.
  • Sakimoto S; Division of Clinical Radiology, Tottori University Hospital.
  • Fujii S; Division of Clinical Radiology, Tottori University Hospital.
  • Ida T; Division of Clinical Radiology, Tottori University Hospital.
  • Moriyama S; Division of Clinical Radiology, Tottori University Hospital.
Nihon Hoshasen Gijutsu Gakkai Zasshi ; 73(10): 1039-1044, 2017.
Article em Ja | MEDLINE | ID: mdl-29057775
ABSTRACT

PURPOSE:

The frame-of-reference using computed-tomography (CT) coordinate system on single-photon emission computed tomography (SPECT) reconstruction is one of the advanced characteristics of the xSPECT reconstruction system. The aim of this study was to reveal the influence of the high-resolution frame-of-reference on the xSPECT reconstruction.

METHODS:

99mTc line-source phantom and National Electrical Manufacturers Association (NEMA) image quality phantom were scanned using the SPECT/CT system. xSPECT reconstructions were performed with the reference CT images in different sizes of the display field-of-view (DFOV) and pixel.

RESULTS:

The pixel sizes of the reconstructed xSPECT images were close to 2.4 mm, which is acquired as originally projection data, even if the reference CT resolution was varied. The full width at half maximum (FWHM) of the line-source, absolute recovery coefficient, and background variability of image quality phantom were independent on the sizes of DFOV in the reference CT images.

CONCLUSION:

The results of this study revealed that the image quality of the reconstructed xSPECT images is not influenced by the resolution of frame-of-reference on SPECT reconstruction.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Tomografia Computadorizada de Emissão de Fóton Único Tipo de estudo: Prognostic_studies Idioma: Ja Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Tomografia Computadorizada de Emissão de Fóton Único Tipo de estudo: Prognostic_studies Idioma: Ja Ano de publicação: 2017 Tipo de documento: Article