Clinical validation of automatic local activation time annotation during focal premature ventricular complex ablation procedures.
Acosta, Juan; Soto-Iglesias, David; Fernández-Armenta, Juan; Frutos-López, Manuel; Jáuregui, Beatriz; Arana-Rueda, Eduardo; Fernández, Marcos; Penela, Diego; Alcaine, Alejandro; Cano, Lucas; Pedrote, Alonso; Berruezo, Antonio.
Afiliação
  • Acosta J; Arrhythmia Section, Department of Cardiology, Virgen del Rocío University Hospital, Avda Manuel Siurot s/n, Seville, Spain.
  • Soto-Iglesias D; Arrhythmia Section, Department of Cardiology, Thorax Institute, Hospital Clínic and IDIBAPS (Institut d'Investigació Agustí Pi i Sunyer), C/Villarroel 170, Barcelona, Catalonia, Spain.
  • Fernández-Armenta J; Arrhythmia Section, Department of Cardiology, Puerta del Mar University Hospital, C/ Ana de Viya 21. Cádiz, Spain.
  • Frutos-López M; Arrhythmia Section, Department of Cardiology, Virgen del Rocío University Hospital, Avda Manuel Siurot s/n, Seville, Spain.
  • Jáuregui B; Arrhythmia Section, Department of Cardiology, Virgen del Rocío University Hospital, Avda Manuel Siurot s/n, Seville, Spain.
  • Arana-Rueda E; Arrhythmia Section, Department of Cardiology, Virgen del Rocío University Hospital, Avda Manuel Siurot s/n, Seville, Spain.
  • Fernández M; Arrhythmia Section, Department of Cardiology, Puerta del Mar University Hospital, C/ Ana de Viya 21. Cádiz, Spain.
  • Penela D; Arrhythmia Section, Department of Cardiology, Thorax Institute, Hospital Clínic and IDIBAPS (Institut d'Investigació Agustí Pi i Sunyer), C/Villarroel 170, Barcelona, Catalonia, Spain.
  • Alcaine A; PhySense Group, Department de Tecnologies de la Informació i les Comunicacions (DTIC), Universitat Pompeu Fabra, Carrer Roc Boronat 138, Tànger Building, Office 55.107, Barcelona, Spain.
  • Cano L; Arrhythmia Section, Department of Cardiology, Puerta del Mar University Hospital, C/ Ana de Viya 21. Cádiz, Spain.
  • Pedrote A; Arrhythmia Section, Department of Cardiology, Virgen del Rocío University Hospital, Avda Manuel Siurot s/n, Seville, Spain.
  • Berruezo A; Arrhythmia Section, Department of Cardiology, Thorax Institute, Hospital Clínic and IDIBAPS (Institut d'Investigació Agustí Pi i Sunyer), C/Villarroel 170, Barcelona, Catalonia, Spain.
Europace ; 20(FI2): f171-f178, 2018 09 01.
Article em En | MEDLINE | ID: mdl-29106546
ABSTRACT

Aims:

Current navigation systems incorporate algorithms for automatic identification of local activation time (LAT). However, data about their utility and accuracy in premature ventricular complex (PVC) ablation procedures are scarce. This study analyses the accuracy of an algorithmic method based on automatic annotation of the maximal negative slope of the unipolar electrogram within the window demarcated by the bipolar electrogram compared with conventional manual annotation during PVC ablation procedures. Methods and

results:

Forty patients with successful ablation of focal PVC in three centres were included. Electroanatomical activation maps obtained with the automatic system (WF-map) were compared with manual annotation maps (M-map). Correlation and concordance of LAT obtained with both methods were assessed at 3536 points. The distance between the earliest activation site (EAS) and the effective radiofrequency application point (e-RFp) were determined in M-map and WF-map. The distance between WF-EAS and M-EAS was assessed. Successful ablation sites included left ventricular outflow tract (LVOT; 55%), right ventricular outflow tract (40%), and tricuspid annulus (5%). Good correlation was observed between the two annotation approaches (r = 0.655; P < 0.0001). Bland-Altman analysis revealed a systematic delayed detection of LAT by WF-map (bias 33.8 ± 30.9 ms), being higher in LVOT than in the right ventricle (42.6 ± 29.2 vs. 27.2 ± 30.5 ms, respectively; P < 0.0001). No difference in EAS-eRFp distance was observed between M-map and WF-map (1.8 ± 2.8 vs. 1.8 ± 3.4 mm, respectively; P = 0.986). The median (interquartile range) distance between WF-EAS and M-EAS was 2.2(0-6) mm.

Conclusion:

Good correlation was found between M-map and WF-map. Local activation time detection was systematically delayed in WF-map, especially in LVOT. Accurate identification of e-RFp was achieved with both annotation approaches.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Processamento de Sinais Assistido por Computador / Potenciais de Ação / Ablação por Cateter / Complexos Ventriculares Prematuros / Técnicas Eletrofisiológicas Cardíacas / Frequência Cardíaca Tipo de estudo: Clinical_trials / Prognostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Processamento de Sinais Assistido por Computador / Potenciais de Ação / Ablação por Cateter / Complexos Ventriculares Prematuros / Técnicas Eletrofisiológicas Cardíacas / Frequência Cardíaca Tipo de estudo: Clinical_trials / Prognostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Ano de publicação: 2018 Tipo de documento: Article