Evaluation of the dialyser inner diameter in online haemodiafiltration. / Valoración del diámetro interno del dializador en hemodiafiltración on-line.
Maduell, Francisco; Ojeda, Raquel; Belmar, Lara; Munguía, Paula; Sango, Cristina; Martinez-Díaz, Ana Isabel; Arias-Guillén, Marta; Vera, Manel; Fontseré, Néstor; Gómez, Miquel; Rodas, Lida; Rico, Nayra.
Afiliação
  • Maduell F; Servicio de Nefrología, Hospital Clínic, Barcelona, España. Electronic address: fmaduell@clinic.ub.es.
  • Ojeda R; Servicio de Nefrología, Hospital Clínic, Barcelona, España.
  • Belmar L; Servicio de Nefrología, Hospital Clínic, Barcelona, España.
  • Munguía P; Servicio de Nefrología, Hospital Clínic, Barcelona, España.
  • Sango C; Servicio de Nefrología, Hospital Clínic, Barcelona, España.
  • Martinez-Díaz AI; Servicio de Nefrología, Hospital Clínic, Barcelona, España.
  • Arias-Guillén M; Servicio de Nefrología, Hospital Clínic, Barcelona, España.
  • Vera M; Servicio de Nefrología, Hospital Clínic, Barcelona, España.
  • Fontseré N; Servicio de Nefrología, Hospital Clínic, Barcelona, España.
  • Gómez M; Servicio de Nefrología, Hospital Clínic, Barcelona, España.
  • Rodas L; Servicio de Nefrología, Hospital Clínic, Barcelona, España.
  • Rico N; Servicio de Bioquímica, Hospital Clínic, Barcelona, España.
Nefrologia (Engl Ed) ; 38(1): 34-40, 2018.
Article em En, Es | MEDLINE | ID: mdl-29325671
ABSTRACT

INTRODUCTION:

Online haemodiafiltration (OL-HDF) has been associated with increased survival. To date, the influence of the inner diameter of the hollow fibres of the dialyser on convective volume has not been well established. The objective of the study was to evaluate the effect of increasing the inner diameter of the dialyser on the convective volume and removal capacity. MATERIAL AND

METHODS:

We included 16 patients in posdilutional OL-HDF with autosubstitution. Each patient was analysed in 4 sessions in which the inner diameter varied; 185µm (FX60 Cordiax and FX80 Cordiax) versus 210µm (FX600 Cordiax and FX800 Cordiax). Different solutes were measured at the beginning and end of each dialysis session.

RESULTS:

No differences in the convective volume were found with an increased inner diameter 32.3±3.1 vs. 31.8±3.6 l/session (FX60 vs. FX600) and 33.7±4.3 vs. 33.5±3.8 l/session (FX80 vs. FX800). The reduction percentages also did not differ urea 83.7±4.5 vs. 84.1±3.4 for FX60 and FX600, and 82.7±4.1 vs. 83.6±3.8 for FX80 vs. FX800; creatinine similar 78.2±5.6 vs. 77.8±4.6 y 77.1±5.4 vs. 78.1±4.9; ß2-microglobulin 82.2±4.3 vs. 82.9±4.2, and 82.9±4.7 vs. 84.0±3.8; myoglobin 71.0±10 vs. 70.2±9 and 72.8±11 vs. 75.0±10; prolactin 70.4±9 vs. 68.1±9, and 72.2±10 vs. 73.4±8.2; and α1-microglobulin 22.9±10 vs. 21.6±10, and 26.5±12 vs. 28.8±11, respectively.

CONCLUSION:

The increase in the inner diameter of the hollow fibres did not result in improved convective volume and removal capacity.