[The Measurement Precision and Accuracy of T1 Mapping Using Polarity Corrected (PC) TI Prep Tool].
Nihon Hoshasen Gijutsu Gakkai Zasshi
; 74(1): 13-21, 2018.
Article
em Ja
| MEDLINE
| ID: mdl-29353832
The aim of this study was to evaluate the measurement precision and accuracy of T1 mapping using a polarity corrected (PC) TI prep tool, which was based on fast field echo (FFE) and obtained one data point with one inversion recovery (IR) pulse. A phantom was used consisting of eight materials with different Gd concentrations. T1 mappings were measured by changing the trigger interval and the inversion time (TI) interval. The T1 mapping measurement precision using the PC TI prep tool increased as the trigger interval was made longer. The measurement precision didn't depend on the interval of TI. On the other hand, when the trigger intervals are more than 1000 ms, the measurement accuracy was less than approximately 8%. By setting the optimal end of TI, the T1 mapping using a PC TI prep tool could measure the T1 value precisely and accurately.