Your browser doesn't support javascript.
loading
Pharmacokinetic and Safety Evaluation of a Transscleral Sustained Unoprostone Release Device in Monkey Eyes.
Nagai, Nobuhiro; Yamada, Shinji; Kawasaki, Junichi; Koyanagi, Eri; Saijo, Saaya; Kaji, Hirokazu; Nishizawa, Matsuhiko; Nakazawa, Toru; Abe, Toshiaki.
Afiliação
  • Nagai N; Division of Clinical Cell Therapy, United Centers for Advanced Research and Translational Medicine (ART), Tohoku University Graduate School of Medicine, Aoba-ku, Sendai, Japan.
  • Yamada S; Division of Clinical Cell Therapy, United Centers for Advanced Research and Translational Medicine (ART), Tohoku University Graduate School of Medicine, Aoba-ku, Sendai, Japan.
  • Kawasaki J; Sucampo Pharma, LLC, Chiyoda-ku, Tokyo, Japan.
  • Koyanagi E; Division of Clinical Cell Therapy, United Centers for Advanced Research and Translational Medicine (ART), Tohoku University Graduate School of Medicine, Aoba-ku, Sendai, Japan.
  • Saijo S; Division of Clinical Cell Therapy, United Centers for Advanced Research and Translational Medicine (ART), Tohoku University Graduate School of Medicine, Aoba-ku, Sendai, Japan.
  • Kaji H; Department of Finemechanics, Graduate School of Engineering, Tohoku University, Aoba-ku, Sendai, Japan.
  • Nishizawa M; Department of Finemechanics, Graduate School of Engineering, Tohoku University, Aoba-ku, Sendai, Japan.
  • Nakazawa T; Department of Ophthalmology, Tohoku University Graduate School of Medicine, Aoba-ku, Sendai, Japan.
  • Abe T; Division of Clinical Cell Therapy, United Centers for Advanced Research and Translational Medicine (ART), Tohoku University Graduate School of Medicine, Aoba-ku, Sendai, Japan.
Invest Ophthalmol Vis Sci ; 59(2): 644-652, 2018 02 01.
Article em En | MEDLINE | ID: mdl-29392308
ABSTRACT

Purpose:

We evaluate the ocular tissue distribution and retinal toxicity of unoprostone (UNO) during 12 months, after transscleral sustained-UNO administration using a drug delivery device in monkey eyes.

Methods:

The device consisted of a reservoir, controlled-release cover, and a drug formulation of photopolymerized polyethylene glycol dimethacrylate. Six mg UNO was loaded into the device (length, 17 mm; width, 4.4 mm; height, 1 mm). The concentrations of M1, a primary metabolite of UNO, in the retina, choroid, vitreous, lens, aqueous humor, iris, ciliary body, and plasma were determined by liquid chromatography-tandem mass spectrometry at 3, 6, and 12 months after implantation. Retinal toxicity was evaluated by electroretinography (ERG), optical coherence tomography (OCT), and IOP at preimplantation, and at 6, 9, and 12 months after implantation. Focal ERGs were performed at 9 and 12 months after implantation.

Results:

M1 was detected in the choroid and retina with maximum peaks of 243.2 and 8.41 ng/g at 6 months, respectively. M1 in the ciliary body and iris was detected with maximum peaks of 7.66 and 10.4 ng/g at 6 and 12 months, respectively. Less than 1 ng/mL or ng/g of M1 was detected in the aqueous humor, vitreous, and lens. No changes were observed in retinal function as assessed by ERG, IOP, or macula thickness and retinal histology by OCT examinations during the 12-month period. No differences in focal ERG amplitudes, especially in the macula, were observed.

Conclusions:

The device provided intraocular sustained delivery of UNO for 12 months without producing severe retinal toxicity.