Beam profile assessment in spectral CT scanners.
Anjomrouz, Marzieh; Shamshad, Muhammad; Panta, Raj K; Broeke, Lieza Vanden; Schleich, Nanette; Atharifard, Ali; Aamir, Raja; Bheesette, Srinidhi; Walsh, Michael F; Goulter, Brian P; Bell, Stephen T; Bateman, Christopher J; Butler, Anthony P H; Butler, Philip H.
Afiliação
  • Anjomrouz M; Department of Radiology, University of Otago, Christchurch, New Zealand.
  • Shamshad M; MARS Bioimaging Ltd, Christchurch, New Zealand.
  • Panta RK; Department of Radiology, University of Otago, Christchurch, New Zealand.
  • Broeke LV; MARS Bioimaging Ltd, Christchurch, New Zealand.
  • Schleich N; Department of Radiology, University of Otago, Christchurch, New Zealand.
  • Atharifard A; MARS Bioimaging Ltd, Christchurch, New Zealand.
  • Aamir R; Department of Physics and Astronomy, University of Canterbury, Christchurch, New Zealand.
  • Bheesette S; Department of Radiation Therapy, University of Otago, Wellington, New Zealand.
  • Walsh MF; Department of Radiology, University of Otago, Christchurch, New Zealand.
  • Goulter BP; MARS Bioimaging Ltd, Christchurch, New Zealand.
  • Bell ST; Department of Radiology, University of Otago, Christchurch, New Zealand.
  • Bateman CJ; MARS Bioimaging Ltd, Christchurch, New Zealand.
  • Butler APH; Department of Radiology, University of Otago, Christchurch, New Zealand.
  • Butler PH; European Organization for Nuclear Research (CERN), Geneva, Switzerland.
J Appl Clin Med Phys ; 19(2): 287-297, 2018 Mar.
Article em En | MEDLINE | ID: mdl-29411524
ABSTRACT
In this paper, we present a method that uses a combination of experimental and modeled data to assess properties of x-ray beam measured using a small-animal spectral scanner. The spatial properties of the beam profile are characterized by beam profile shape, the angular offset along the rotational axis, and the photon count difference between experimental and modeled data at the central beam axis. Temporal stability of the beam profile is assessed by measuring intra- and interscan count variations. The beam profile assessment method was evaluated on several spectral CT scanners equipped with Medipix3RX-based detectors. On a well-calibrated spectral CT scanner, we measured an integral count error of 0.5%, intrascan count variation of 0.1%, and an interscan count variation of less than 1%. The angular offset of the beam center ranged from 0.8° to 1.6° for the studied spectral CT scanners. We also demonstrate the capability of this method to identify poor performance of the system through analyzing the deviation of the experimental beam profile from the model. This technique can, therefore, aid in monitoring the system performance to obtain a robust spectral CT; providing the reliable quantitative images. Furthermore, the accurate offset parameters of a spectral scanner provided by this method allow us to incorporate a more realistic form of the photon distribution in the polychromatic-based image reconstruction models. Both improvements of the reliability of the system and accuracy of the volume reconstruction result in a better discrimination and quantification of the imaged materials.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Processamento de Imagem Assistida por Computador / Tomografia Computadorizada por Raios X / Tomógrafos Computadorizados / Modelos Teóricos Limite: Humans Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Processamento de Imagem Assistida por Computador / Tomografia Computadorizada por Raios X / Tomógrafos Computadorizados / Modelos Teóricos Limite: Humans Idioma: En Ano de publicação: 2018 Tipo de documento: Article