Micro-computed tomography for the quantification of blocked fibers in hemodialyzers.
Vanommeslaeghe, Floris; Van Biesen, Wim; Dierick, Manuel; Boone, Matthieu; Dhondt, Annemieke; Eloot, Sunny.
Afiliação
  • Vanommeslaeghe F; Nephrology Department, Ghent University Hospital, Ghent, Belgium. Floris.Vanommeslaeghe@ugent.be.
  • Van Biesen W; Nephrology Department, Ghent University Hospital, Ghent, Belgium.
  • Dierick M; UGCT, Department of Subatomic and Radiation Physics, Ghent University, Ghent, Belgium.
  • Boone M; UGCT, Department of Subatomic and Radiation Physics, Ghent University, Ghent, Belgium.
  • Dhondt A; Nephrology Department, Ghent University Hospital, Ghent, Belgium.
  • Eloot S; Nephrology Department, Ghent University Hospital, Ghent, Belgium.
Sci Rep ; 8(1): 2677, 2018 02 08.
Article em En | MEDLINE | ID: mdl-29422614
ABSTRACT
A novel technique based on micro-CT scanning is developed to quantify coagulation in fibers of hemodialyzers. This objectivation is needed to allow accurate assessment of thrombogenicity of dialyzers used during hemodialysis, for example when comparing different strategies to avoid coagulation and/or fiber blocking. The protocol allowed imaging at a resolution of 25 µm, making it possible to count the open, non-coagulated fibers in a non-invasive way. In 3 fresh, non-used FX600 hemodialyzers, patent fiber counts were extremely consistent (10748 ± 2). To illustrate the potential of this technique, different dialysis parameters currently used as surrogates for fiber blocking were evaluated during 20 hemodialysis sessions. After dialysis, the FX600 dialyzers were visually scored for clotting, dried and subsequently weighed and scanned. The number of patent fibers (10003 [8763,10330], range 534-10692) did not correlate with any of the recorded surrogate parameters. Micro-CT scanning is a feasible, objective, non-invasive, accurate and reproducible tool for quantification of the degree of fiber blocking in a hemodialyzer after use, making it a potential gold standard for use in studies on fiber blocking during renal replacement therapies.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Diálise Renal / Microtomografia por Raio-X / Rins Artificiais Tipo de estudo: Guideline / Observational_studies Limite: Humans Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Diálise Renal / Microtomografia por Raio-X / Rins Artificiais Tipo de estudo: Guideline / Observational_studies Limite: Humans Idioma: En Ano de publicação: 2018 Tipo de documento: Article