OBJECTIVE: To determine the circuit elements required to theoretically describe the stimulus waveforms generated by an implantable pulse generator (IPG) during clinical deep brain stimulation (DBS). METHODS: We experimentally interrogated the Medtronic Activa PC DBS IPG and defined an equivalent circuit model that accurately captured the output of the IPG. We then compared the detailed circuit model of the clinical stimulus waveforms to simplified representations commonly used in computational models of DBS. We quantified the errors associated with these simplifications using theoretical activation thresholds of myelinated axons in response to DBS. RESULTS: We found that the detailed IPG model generated substantial differences in activation thresholds compared to simplified models. These differences were largest for bipolar stimulation with long pulse widths. Average errors were â¼3 to 24% for voltage-controlled stimulation and â¼2 to 11% for current-controlled stimulation. CONCLUSIONS: Our results demonstrate the importance of including basic circuit elements (e.g. blocking capacitors, lead wire resistance, electrode capacitance) in model analysis of DBS. SIGNIFICANCE: Computational models of DBS are now commonly used in academic research, industrial technology development, and in the selection of clinical stimulation parameters. Incorporating a realistic representation of the IPG output is necessary to improve the accuracy and utility of these clinical and scientific tools.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Simulação por Computador
/
Redes Neurais de Computação
/
Estimulação Encefálica Profunda
/
Eletrodos Implantados
Limite:
Humans
Idioma:
En
Ano de publicação:
2018
Tipo de documento:
Article