Magnetic resonance imaging safety in nonconditional pacemaker and defibrillator recipients: A meta-analysis and systematic review.
Shah, Anand D; Morris, Mike A; Hirsh, David S; Warnock, Megan; Huang, Yijian; Mollerus, Michael; Merchant, Faisal M; Patel, Anshul M; Delurgio, David B; Patel, Adarsh U; Hoskins, Michael H; El Chami, Mikhael F; Leon, Angel R; Langberg, Jonathan J; Lloyd, Michael S.
Afiliação
  • Shah AD; Department of Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, Georgia. Electronic address: adshah7@emory.edu.
  • Morris MA; Department of Medicine, St. George's University School of Medicine, Grenada, West Indies.
  • Hirsh DS; Department of Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, Georgia.
  • Warnock M; Emory University, Rollins School of Biostatistics and Bioinformatics, Atlanta, Georgia.
  • Huang Y; Emory University, Rollins School of Biostatistics and Bioinformatics, Atlanta, Georgia.
  • Mollerus M; Department of Cardiology, Essentia Health, Duluth, Minnesota.
  • Merchant FM; Department of Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, Georgia.
  • Patel AM; Department of Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, Georgia.
  • Delurgio DB; Department of Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, Georgia.
  • Patel AU; Department of Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, Georgia.
  • Hoskins MH; Department of Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, Georgia.
  • El Chami MF; Department of Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, Georgia.
  • Leon AR; Department of Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, Georgia.
  • Langberg JJ; Department of Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, Georgia.
  • Lloyd MS; Department of Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, Georgia.
Heart Rhythm ; 15(7): 1001-1008, 2018 07.
Article em En | MEDLINE | ID: mdl-29458192
ABSTRACT

BACKGROUND:

Recommendations regarding performance of magnetic resonance imaging (MRI) in non-MRI conditional pacemaker and defibrillator recipients are evolving. Previous studies have suggested low adverse event rates with MRI in nonconditional cardiac implantable electronic device (CIED) recipients, but low power limits optimal characterization of risk.

OBJECTIVE:

The purpose of this study was to perform a systematic review and meta-analysis to characterize the clinical risk associated with MRI in CIED recipients in order to improve power.

METHODS:

PubMed and CINAHL indexed articles from 1990 to 2017 were queried. A random effects model was used for meta-analysis of continuous variables. Safety outcomes were evaluated with descriptive statistics.

RESULTS:

Seventy studies of non-MRI conditional devices undergoing MRI were identified, allowing for analysis of 5099 patients who underwent a total of 5908 MRI studies. Heterogeneity in lead parameter changes was observed within studies, although smaller variances were noted between studies. All lead characteristics and battery voltages showed very small, clinically insignificant changes when assessed as a pooled cohort, although cases of clinically relevant outcomes were also noted (lead failure 3, implantable cardioverter-defibrillator shock 1, electrical reset 94). Electrical resets were found only in older devices. Defibrillator function was unchanged, and inappropriate shocks were avoided with pre-MRI programming changes.

CONCLUSION:

This review demonstrated low lead failure and clinical event rates in non-MRI conditional pacemaker and defibrillator recipients undergoing MRI. Observed changes were small and interstudy variance was low, suggesting that the composite event rates offer a reasonable estimate of true effect. The observed adverse events reinforce the need for ongoing vigilance and caution, particularly with older devices.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Arritmias Cardíacas / Desfibriladores Implantáveis / Imagem Cinética por Ressonância Magnética Tipo de estudo: Diagnostic_studies / Guideline / Prognostic_studies / Systematic_reviews Limite: Humans Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Arritmias Cardíacas / Desfibriladores Implantáveis / Imagem Cinética por Ressonância Magnética Tipo de estudo: Diagnostic_studies / Guideline / Prognostic_studies / Systematic_reviews Limite: Humans Idioma: En Ano de publicação: 2018 Tipo de documento: Article