Highly Sensitive and Exceptionally Wide Dynamic Range Detection of Ammonia Gas by Indium Hexacyanoferrate Nanoparticles Using FTIR Spectroscopy.
Manakasettharn, Supone; Takahashi, Akira; Kawamoto, Tohru; Noda, Keiko; Sugiyama, Yutaka; Nakamura, Tohru.
Afiliação
  • Manakasettharn S; Nanomaterials Research Institute, National Institute of Advanced Industrial Science and Technology (AIST) , 1-1-1 Higashi , Tsukuba 305-8565 , Japan.
  • Takahashi A; National Nanotechnology Center (NANOTEC), National Science and Technology Development Agency (NSTDA) , 111 Thailand Science Park, Phahonyothin Road , Khlong Nueng, Khlong Luang , Pathum Thani 12120 , Thailand.
  • Kawamoto T; Nanomaterials Research Institute, National Institute of Advanced Industrial Science and Technology (AIST) , 1-1-1 Higashi , Tsukuba 305-8565 , Japan.
  • Noda K; Nanomaterials Research Institute, National Institute of Advanced Industrial Science and Technology (AIST) , 1-1-1 Higashi , Tsukuba 305-8565 , Japan.
  • Sugiyama Y; Nanomaterials Research Institute, National Institute of Advanced Industrial Science and Technology (AIST) , 1-1-1 Higashi , Tsukuba 305-8565 , Japan.
  • Nakamura T; Nanomaterials Research Institute, National Institute of Advanced Industrial Science and Technology (AIST) , 1-1-1 Higashi , Tsukuba 305-8565 , Japan.
Anal Chem ; 90(7): 4856-4862, 2018 04 03.
Article em En | MEDLINE | ID: mdl-29554801
Ammonia gas is useful but caustic; thus, its concentration is monitored depending on applications. We prepared indium hexacyanoferrate nanoparticles (InHCF-NPs, HCF = [FeII(CN)6]4-) with average diameter around 8 nm by simple reaction at room temperature between In cations and HCF anions, and we found the unique functionality of InHCF-NPs which were capable of highly sensitive (16 ppb) and exceptionally wide range (190000, 16 ppb -0.3%) detection of ammonia gas within 6 min by IR measurements. Slope changes of the IR peak ratio of adsorbed ammonium over CN moieties in the InHCF framework indicated a good log-log linear correlation along gas concentrations, in which a wide dynamic range over 105 was realized for the first time in the field of ammonia gas detection.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2018 Tipo de documento: Article