Your browser doesn't support javascript.
loading
Myocardial perfusion quantification using simultaneously acquired 13 NH3 -ammonia PET and dynamic contrast-enhanced MRI in patients at rest and stress.
Kunze, Karl P; Nekolla, Stephan G; Rischpler, Christoph; Zhang, Shelley HuaLei; Hayes, Carmel; Langwieser, Nicolas; Ibrahim, Tareq; Laugwitz, Karl-Ludwig; Schwaiger, Markus.
Afiliação
  • Kunze KP; Klinikum rechts der Isar der TU München, Department of Nuclear Medicine, Munich, Germany.
  • Nekolla SG; Klinikum rechts der Isar der TU München, Department of Nuclear Medicine, Munich, Germany.
  • Rischpler C; DZHK (Deutsches Zentrum für Herz-Kreislauf-Forschung e.V.) partner site Munich Heart Alliance, Munich, Germany.
  • Zhang SH; Klinikum rechts der Isar der TU München, Department of Nuclear Medicine, Munich, Germany.
  • Hayes C; DZHK (Deutsches Zentrum für Herz-Kreislauf-Forschung e.V.) partner site Munich Heart Alliance, Munich, Germany.
  • Langwieser N; Brigham and Women's Hospital, Department of Radiology, Boston, United States.
  • Ibrahim T; Siemens Healthcare GmbH, Erlangen, Germany.
  • Laugwitz KL; DZHK (Deutsches Zentrum für Herz-Kreislauf-Forschung e.V.) partner site Munich Heart Alliance, Munich, Germany.
  • Schwaiger M; Klinikum rechts der Isar der TU München, Department of Cardiology, Munich, Germany.
Magn Reson Med ; 80(6): 2641-2654, 2018 12.
Article em En | MEDLINE | ID: mdl-29672922
ABSTRACT

PURPOSE:

Systematic differences with respect to myocardial perfusion quantification exist between DCE-MRI and PET. Using the potential of integrated PET/MRI, this study was conceived to compare perfusion quantification on the basis of simultaneously acquired 13 NH3 -ammonia PET and DCE-MRI data in patients at rest and stress.

METHODS:

Twenty-nine patients were examined on a 3T PET/MRI scanner. DCE-MRI was implemented in dual-sequence design and additional T1 mapping for signal normalization. Four different deconvolution methods including a modified version of the Fermi technique were compared against 13 NH3 -ammonia results.

RESULTS:

Cohort-average flow comparison yielded higher resting flows for DCE-MRI than for PET and, therefore, significantly lower DCE-MRI perfusion ratios under the common assumption of equal arterial and tissue hematocrit. Absolute flow values were strongly correlated in both slice-average (R2 = 0.82) and regional (R2 = 0.7) evaluations. Different DCE-MRI deconvolution methods yielded similar flow result with exception of an unconstrained Fermi method exhibiting outliers at high flows when compared with PET.

CONCLUSION:

Thresholds for Ischemia classification may not be directly tradable between PET and MRI flow values. Differences in perfusion ratios between PET and DCE-MRI may be lifted by using stress/rest-specific hematocrit conversion. Proper physiological constraints are advised in model-constrained deconvolution.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Tomografia por Emissão de Pósitrons / Imagem de Perfusão do Miocárdio Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Tomografia por Emissão de Pósitrons / Imagem de Perfusão do Miocárdio Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article