Evaluation of an accelerated 3D SPACE sequence with compressed sensing and free-stop scan mode for imaging of the knee.
Henninger, B; Raithel, E; Kranewitter, C; Steurer, M; Jaschke, W; Kremser, C.
Afiliação
  • Henninger B; Department of Radiology, Medical University of Innsbruck, Anichstraße 35, Innsbruck, Austria. Electronic address: benjamin.henninger@i-med.ac.at.
  • Raithel E; Siemens Healthcare, Allee am Roethelheimpark 2, Erlangen, Germany.
  • Kranewitter C; Department of Radiology, Medical University of Innsbruck, Anichstraße 35, Innsbruck, Austria.
  • Steurer M; Department of Radiology, Medical University of Innsbruck, Anichstraße 35, Innsbruck, Austria.
  • Jaschke W; Department of Radiology, Medical University of Innsbruck, Anichstraße 35, Innsbruck, Austria.
  • Kremser C; Department of Radiology, Medical University of Innsbruck, Anichstraße 35, Innsbruck, Austria.
Eur J Radiol ; 102: 74-82, 2018 May.
Article em En | MEDLINE | ID: mdl-29685548
ABSTRACT

PURPOSE:

To prospectively evaluate a prototypical 3D turbo-spin-echo proton-density-weighted sequence with compressed sensing and free-stop scan mode for preventing motion artefacts (3D-PD-CS-SPACE free-stop) for knee imaging in a clinical setting. METHODS AND MATERIALS 80 patients underwent 3T magnetic resonance imaging (MRI) of the knee with our 2D routine protocol and with 3D-PD-CS-SPACE free-stop. In case of a scan-stop caused by motion (images are calculated nevertheless) the sequence was repeated without free-stop mode. All scans were evaluated by 2 radiologists concerning image quality of the 3D-PD-CS-SPACE (with and without free-stop). Important knee structures were further assessed in a lesion based analysis and compared to our reference 2D-PD-fs sequences.

RESULTS:

Image quality of the 3D-PD-CS-SPACE free-stop was found optimal in 47/80, slightly compromised in 21/80, moderately in 10/80 and severely in 2/80. In 29/80, the free-stop scan mode stopped the 3D-PD-CS-SPACE due to subject motion with a slight increase of image quality at longer effective acquisition times. Compared to the 3D-PD-CS-SPACE with free-stop, the image quality of the acquired 3D-PD-CS-SPACE without free-stop was found equal in 6/29, slightly improved in 13/29, improved with equal contours in 8/29, and improved with sharper contours in 2/29. The lesion based analysis showed a high agreement between the results from the 3D-PD-CS-SPACE free-stop and our 2D-PD-fs routine protocol (overall agreement 96.25%-100%, Cohen's Kappa 0.883-1, p < 0.001).

CONCLUSION:

3D-PD-CS-SPACE free-stop is a reliable alternative for standard 2D-PD-fs protocols with acceptable acquisition times.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética / Imageamento Tridimensional / Articulação do Joelho Tipo de estudo: Guideline / Observational_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética / Imageamento Tridimensional / Articulação do Joelho Tipo de estudo: Guideline / Observational_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article