Your browser doesn't support javascript.
loading
Critical-depth Raman spectroscopy enables home-use non-invasive glucose monitoring.
Lundsgaard-Nielsen, Signe M; Pors, Anders; Banke, Stefan O; Henriksen, Jan E; Hepp, Dietrich K; Weber, Anders.
Afiliação
  • Lundsgaard-Nielsen SM; RSP Systems, Odense, Denmark.
  • Pors A; RSP Systems, Odense, Denmark.
  • Banke SO; RSP Systems, Odense, Denmark.
  • Henriksen JE; Department of Endocrinology, Odense University Hospital, Odense, Denmark.
  • Hepp DK; Endocrinology and Diabetology, Munich, Germany.
  • Weber A; RSP Systems, Odense, Denmark.
PLoS One ; 13(5): e0197134, 2018.
Article em En | MEDLINE | ID: mdl-29750797
ABSTRACT
One of the most ambitious endeavors in the field of diabetes technology is non-invasive glucose sensing. In the past decades, a number of different technologies have been assessed, but none of these have found its entry into general clinical use. We report on the development of a table-top confocal Raman spectrometer that was used in the home of patients with diabetes and operated for extended periods of time unsupervised and without recalibration. The system is based on measurement of glucose levels at a 'critical depth' in the skin, specifically in the interstitial fluid located below the stratum corneum but above the underlying adipose tissue layer. The region chosen for routine glucose measurements was the base of the thumb (the thenar). In a small clinical study, 35 patients with diabetes analyzed their interstitial fluid glucose for a period of 60 days using the new critical-depth Raman (CD-Raman) method and levels were correlated to reference capillary blood glucose values using a standard finger-stick and test strip product. The calibration of the CD-Raman system was stable for > 10 days. Measurement performance for glucose levels present at, or below, a depth of ~250µm below the skin surface was comparable to that reported for currently available invasive continuous glucose monitors. In summary, using the CD-Raman technology we have demonstrated the first successful use of a non-invasive glucose monitor in the home.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Análise Espectral Raman / Glicemia / Diabetes Mellitus Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Análise Espectral Raman / Glicemia / Diabetes Mellitus Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article