Cusp Tear of Trifecta Aortic Bioprosthesis Resulting in Acute Heart Failure.
J Heart Valve Dis
; 26(5): 592-594, 2017 09.
Article
em En
| MEDLINE
| ID: mdl-29762930
Herein is presented the case of an 83-year-old male patient in cardiogenic shock with acute aortic regurgitation that occurred six years after aortic valve replacement (AVR) with a 23 mm Trifecta™ valve. Prosthesis endocarditis was initially suspected because of a floating structure attached to the aortic valve that was visible on echocardiography. Emergency redo-AVR surgery was performed, but no signs of endocarditis were found intraoperatively. Hence, cusp tearing of the implanted bioprosthesis was considered to be the reason for the severe aortic regurgitation.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Insuficiência da Valva Aórtica
/
Estenose da Valva Aórtica
/
Complicações Pós-Operatórias
/
Reoperação
/
Choque Cardiogênico
/
Bioprótese
/
Falha de Prótese
/
Próteses Valvulares Cardíacas
/
Implante de Prótese de Valva Cardíaca
/
Efeitos Adversos de Longa Duração
Tipo de estudo:
Diagnostic_studies
/
Etiology_studies
Limite:
Aged80
/
Humans
/
Male
Idioma:
En
Ano de publicação:
2017
Tipo de documento:
Article