Robust computation in 2D absolute EIT (a-EIT) using D-bar methods with the 'exp' approximation.
Physiol Meas
; 39(6): 064005, 2018 06 27.
Article
em En
| MEDLINE
| ID: mdl-29846182
ABSTRACT
OBJECTIVE:
Absolute images have important applications in medical electrical impedance tomography (EIT) imaging, but the traditional minimization and statistical based computations are very sensitive to modeling errors and noise. In this paper, it is demonstrated that D-bar reconstruction methods for absolute EIT are robust to such errors.APPROACH:
The effects of errors in domain shape and electrode placement on absolute images computed with 2D D-bar reconstruction algorithms are studied on experimental data. MAINRESULTS:
It is demonstrated with tank data from several EIT systems that these methods are quite robust to such modeling errors, and furthermore the artefacts arising from such modeling errors are similar to those occurring in classic time-difference EIT imaging.SIGNIFICANCE:
This study is promising for clinical applications where absolute EIT images are desirable but previously thought impossible.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Processamento de Imagem Assistida por Computador
/
Tomografia
Idioma:
En
Ano de publicação:
2018
Tipo de documento:
Article