Utilizing a Mini-Humidifier To Deposit Matrix for MALDI Imaging.
Huang, Xi; Zhan, Lingpeng; Sun, Jie; Xue, Jinjuan; Liu, Huihui; Xiong, Caiqiao; Nie, Zongxiu.
Afiliação
  • Huang X; Beijing National Laboratory for Molecular Sciences, Key Laboratory of Analytical Chemistry for Living Biosystems, Institute of Chemistry , Chinese Academy of Sciences , Beijing 100190 , China.
  • Zhan L; University of Chinese Academy of Sciences , Beijing 100049 , China.
  • Sun J; Beijing National Laboratory for Molecular Sciences, Key Laboratory of Analytical Chemistry for Living Biosystems, Institute of Chemistry , Chinese Academy of Sciences , Beijing 100190 , China.
  • Xue J; University of Chinese Academy of Sciences , Beijing 100049 , China.
  • Liu H; Beijing National Laboratory for Molecular Sciences, Key Laboratory of Analytical Chemistry for Living Biosystems, Institute of Chemistry , Chinese Academy of Sciences , Beijing 100190 , China.
  • Xiong C; University of Chinese Academy of Sciences , Beijing 100049 , China.
  • Nie Z; Beijing National Laboratory for Molecular Sciences, Key Laboratory of Analytical Chemistry for Living Biosystems, Institute of Chemistry , Chinese Academy of Sciences , Beijing 100190 , China.
Anal Chem ; 90(14): 8309-8313, 2018 07 17.
Article em En | MEDLINE | ID: mdl-29939716
ABSTRACT
MALDI mass spectrometry imaging (MALDI-MSI) is a powerful tool to study endogenous metabolites. The process of matrix deposition is crucial for a high-quality imaging result. Commercial instruments for matrix deposition are expensive. Low-cost methods like airbrushing will generate matrix crystals that are too large for high-spatial-resolution imaging. Sublimation may cause some compounds to go undetected because of the lack of solvent. Herein, we utilized a mini-humidifier, costing less than 5 dollars, to deposit matrix for MALDI-MSI. Compared with Imageprep, a commercialized instrument, our device based on the humidifier provided higher sensitivity and much smaller matrix crystals with diameters of less than 10 µm. High-quality ion images with 10 µm spatial resolution were obtained using our method. The enhancement of sensitivity by the humidifier could provide a sufficient amount of ions to perform tandem mass imaging. We also performed MALDI-MS/MS imaging to separate two lipids in mouse brain.