Effect of varying skin surface electrode position on electroretinogram responses recorded using a handheld stimulating and recording system.
Hobby, Angharad E; Kozareva, Diana; Yonova-Doing, Ekaterina; Hossain, Ibtesham T; Katta, Mohamed; Huntjens, Byki; Hammond, Christopher J; Binns, Alison M; Mahroo, Omar A.
Afiliação
  • Hobby AE; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.
  • Kozareva D; Centre of Applied Vision Research, City, University of London, Northampton Square, London, EC1V 0HB, UK.
  • Yonova-Doing E; Department of Twin Research and Genetic Epidemiology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.
  • Hossain IT; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.
  • Katta M; Department of Twin Research and Genetic Epidemiology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.
  • Huntjens B; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.
  • Hammond CJ; Department of Twin Research and Genetic Epidemiology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.
  • Binns AM; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.
  • Mahroo OA; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.
Doc Ophthalmol ; 137(2): 79-86, 2018 10.
Article em En | MEDLINE | ID: mdl-30046929
ABSTRACT

PURPOSE:

A handheld device (the RETeval system, LKC Technologies) aims to increase the ease of electroretinogram (ERG) recording by using specially designed skin electrodes, rather than corneal electrodes. We explored effects of electrode position on response parameters recorded using this device.

METHODS:

Healthy adult twins were recruited from the TwinsUK cohort and underwent recording of light-adapted flicker ERGs (corresponding to international standard stimuli). In Group 1, skin electrodes were placed in a "comfortable" position, which was up to 20 mm below the lid margin. For subsequent participants (Group 2), the electrode was positioned 2 mm from the lid margin as recommended by the manufacturer. Amplitudes and peak times (averaged from both eyes) were compared between groups after age-matching and inclusion of only one twin per pair. Light-adapted flicker and flash ERGs were recorded for an additional 10 healthy subjects in two consecutive recording sessions in the test eye, electrode position was varied from 2 to 10-20 mm below the lid margin between sessions; in the fellow (control) eye, the electrode was 2 mm below the lid margin throughout. Amplitudes and peak times (test eye normalised to control eye) were compared for the two sessions.

RESULTS:

Including one twin per pair, and age-matching yielded 28 individuals per group. Flicker ERG amplitudes were significantly lower for Group 1 than Group 2 participants (p = 0.0024). However, mean peak times did not differ between groups (p = 0.54). For the subjects in whom electrode position was changed between recording sessions, flash and flicker amplitudes were significantly lower when positioned further from the lid margin (p < 0.005), but peak times were similar (p > 0.5).

CONCLUSIONS:

Moving the skin electrodes further from the lid margin significantly reduces response amplitudes, highlighting the importance of consistent electrode positioning. However, this does not significantly affect peak times. Thus, it may be feasible to adopt a more comfortable position in participants who cannot tolerate the recommended position if analysis is restricted to peak time parameters.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sistemas Automatizados de Assistência Junto ao Leito / Eletrorretinografia / Pálpebras Tipo de estudo: Observational_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sistemas Automatizados de Assistência Junto ao Leito / Eletrorretinografia / Pálpebras Tipo de estudo: Observational_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article