Analysis of the long-term stability of the output of electron beams generated by the Novac11™ IORT accelerator.
Bijok, Michal; Gruszczynska, Ewelina; Kowalczyk, Adam; Sikorska, Katarzyna; Walewska, Agnieszka; Kukolowicz, Pawel F.
Afiliação
  • Bijok M; Medical Physics Department, Maria Sklodowska-Curie Memorial Centre and Institute of Oncology, Roentgen Str. 5, 02-781 Warsaw, Poland.
  • Gruszczynska E; Medical Physics Department, Maria Sklodowska-Curie Memorial Centre and Institute of Oncology, Roentgen Str. 5, 02-781 Warsaw, Poland.
  • Kowalczyk A; Medical Physics Department, Maria Sklodowska-Curie Memorial Centre and Institute of Oncology, Roentgen Str. 5, 02-781 Warsaw, Poland.
  • Sikorska K; Medical Physics Department, Maria Sklodowska-Curie Memorial Centre and Institute of Oncology, Roentgen Str. 5, 02-781 Warsaw, Poland.
  • Walewska A; Medical Physics Department, Maria Sklodowska-Curie Memorial Centre and Institute of Oncology, Roentgen Str. 5, 02-781 Warsaw, Poland.
  • Kukolowicz PF; Medical Physics Department, Maria Sklodowska-Curie Memorial Centre and Institute of Oncology, Roentgen Str. 5, 02-781 Warsaw, Poland.
Rep Pract Oncol Radiother ; 23(5): 341-345, 2018.
Article em En | MEDLINE | ID: mdl-30127674
ABSTRACT

AIM:

The aim of the study was to analyze the long-term stability of electron beams generated by the Novac11™ IORT accelerator.

BACKGROUND:

Novac11™ (NRT®) is a mobile electron accelerator designed to irradiate small areas of tissue, up to 10 cm in diameter, with electron beams during surgical procedures. It is characterized by a great mobility guaranteed by a number of degrees of freedom enabling irradiation in the conditions of an operating theatre. MATERIALS AND

METHODS:

Over the period of January 2013 and September 2016, the measurement sessions of the output of clinically used beam qualities (6, 8 and 10 MeV) were carried out 41 times. Because of the unsatisfactory long-term stability, an extra procedure of tuning of the magnetron, suggested by the manufacturer, was introduced in October 2015, 15 measurements were performed since then. The output of the Novac11™ accelerator was measured in the reference conditions recommended by the IAEA Report 398, the measurements of the charge in the ionization chamber at the reference depth were carried out with a Dose1™ electrometer and a plane-parallel chamber PPC05™ from IBA®.

RESULTS:

The introduction of the tuning of the magnetron procedure resulted in satisfactory long-term stability of the measured outputs below 2%.

CONCLUSIONS:

After the introduction of the STV parameter tuning procedure, the long-term stability of the Novac11™ output increased considerably and is within the values declared by the manufacturer.